The Effect of Vascular Access Location and Size on Circuit Survival in Pediatric Continuous Renal Replacement Therapy: A Report from the PPCRRT Registry
Authors:
- 1 December 2007
- journal article
- research article
- Published by SAGE Publications in The International Journal of Artificial Organs
- Vol. 30 (12) , 1116-1121
Abstract
Purpose: Well-functioning vascular access is essential for the provision of adequate CRRT However, few data exist to describe the effect of catheter size or location on CRRT performance in the pediatric population. Methods: Data for vascular access site, size, and location, as well as type of anticoagulant used and patient demographic data were gathered from the ppCRRT registry. Kaplan-Meier curves were generated and then analyzed by log-rank test or Cox Proportional Hazards model. Results: Access diameter was found to significantly affect circuit survival. None of the 5 French catheters lasted longer than 20 hours. Seven and 9 French, but not 8 French, catheters fared worse than larger diameter catheters (p=0.002). Circuits associated with internal jugular access survived longer than subclavian or femoral access associated circuits (p<0.05). Circuit survival was also found to be favorably associated with the CVVHD modality (p<0.001). Conclusions: Functional CRRT circuit survival in children is favored by larger catheter diameter, internal jugular vein insertion site and CVVHD. For patients requiring catheter diameters less than 10 French, CRRT circuit survival might be optimized if internal jugular vein insertion is feasible. Conversely, when a vascular access site other than the internal jugular vein is most prudent, consideration should be given to using the largest diameter catheter appropriate for the size of the child. The CVVHD modality was associated with longer circuit survival, but the mechanism by which this occurs is unclear.Keywords
This publication has 10 references indexed in Scilit:
- Solute removal during continuous renal replacement therapy in critically ill patients: convection versus diffusionCritical Care, 2006
- Multi-centre evaluation of anticoagulation in patients receiving continuous renal replacement therapy (CRRT)Nephrology Dialysis Transplantation, 2005
- The Prospective Pediatric Continuous Renal Replacement Therapy (ppCRRT) Registry: Design, Development and Data AssessedThe International Journal of Artificial Organs, 2004
- Relationship between Blood Flow, Access Catheter and Circuit Failure during CRRT: A Practical Review2004
- The Acute Dialysis Quality Initiative—Part VI: Access and anticoagulation in CRRTAdvances in Renal Replacement Therapy, 2002
- Pediatric hemofiltration: Normocarb dialysate solution with citrate anticoagulationPediatric Nephrology, 2002
- Catheter PerformanceSeminars in Dialysis, 2001
- Access recirculation in temporary hemodialysis catheters as measured by the saline dilution techniqueAmerican Journal of Kidney Diseases, 2000
- Continuous venovenous hemodiafiltration in infants and childrenAmerican Journal of Kidney Diseases, 1995
- Effects of Access Catheter Dimensions on Bloodflow in Continuous Arteriovenous Hemofiltration11991