Characterization of blood mimicking fluid with ultrafast ultrasonic and optical image velocimeters
- 31 May 2022
- journal article
- research article
- Published by IOP Publishing in Japanese Journal of Applied Physics
- Vol. 61 (SG) , SG1067
Abstract
A blood mimicking fluid (BMF) is imperative for the evaluation of Doppler ultrasound. Doppler ultrasound still causes errors due to some artifacts such as aliasing and presence of grating lobes. One of the other velocimeters is the optical particle image velocimeter (PIV). This study initially developed an in vitro measurement system for analyzing flowing BMF with ultrasonic and optical PIVs. The acoustic properties such as speed of sound, attenuation, and backscatter coefficient of BMF equivalent to the human blood, used for both ultrasonic and optical PIVs were analyzed in a frequency range of 4–12 MHz. The velocity profiles were estimated by ultrasonic and optical PIVs using a block matching method. A difference between velocities obtained by ultrasonic and optical data was within 4.0% using BMF with 20 μ m polyamide particle at 0.2% concentration that realized the acoustic properties and speckle patterns similar to those in ultrafast ultrasound blood flow imaging.Keywords
This publication has 37 references indexed in Scilit:
- Profound Increase in Longitudinal Displacements of the Porcine Carotid Artery Wall Can Take Place Independently of Wall Shear Stress: A Continuation ReportUltrasound in Medicine & Biology, 2015
- Experimental evaluation of spectral-based quantitative ultrasound imaging using plane wave compoundingIEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 2014
- On the estimation of backscatter coefficients using single-element focused transducersThe Journal of the Acoustical Society of America, 2011
- In Vitro and Preliminary In Vivo Validation of Echo Particle Image Velocimetry in Carotid Vascular ImagingUltrasound in Medicine & Biology, 2011
- Extended three-dimensional impedance map methods for identifying ultrasonic scattering sitesThe Journal of the Acoustical Society of America, 2008
- Longitudinal movements and resulting shear strain of the arterial wallAmerican Journal of Physiology-Heart and Circulatory Physiology, 2006
- Development and validation of echo PIVExperiments in Fluids, 2004
- Methods for Estimation of Subsample Time Delays of Digitized Echo SignalsUltrasonic Imaging, 1995
- Backscatter Coefficient Measurements Using a Reference Phantom to Extract Depth-Dependent Instrumentation FactorsUltrasonic Imaging, 1990
- Speed of sound in water by a direct methodJournal of Research of the National Bureau of Standards, 1957