Individualized gating windows based on four-dimensional CT information for respiration-gated radiotherapy
Authors:
- Ming ChaoMCMing Chao
- Jiji AntonyJAJiji Antony
- Lei XingLXLei Xing
- 19 December 2007
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 53 (1) , 165-175
Abstract
The purpose of this work is to relate the gating window and displacement of a moving tumor target and develop a systematic method to individualize the gating window for respiration-gated radiation therapy (RT). As the relationship between patient anatomy and respiration phase is contained in 4D images, we aim to quantify this information and utilize the data to guide gated treatment planning. After 4D image acquisition, the target and organs at risk were delineated manually on the selected gating phase. The contours were propagated automatically onto every phase-specific image set using a control volume-based contour mapping technique. The mean and maximum distances between the contours in the gating phase and each of other phases were evaluated in three dimensions. The gating window was determined in such a way that the residual movement of the target within the window is smaller or equal to the patient's setup error. The proposed method was applied to plan the gated treatments of 12 lung cancer patients. As a result of this work, a method to calculate patient-specific gating windows has been developed. The general reference drawn from this study is that, with the aide of 4D images and automated 4D contour propagation, it is feasible to individualize the gating widow selection. As compared with the current practice, the proposed technique has a potential to eliminate the guesswork involved in choosing a gating window and avoid dosimetric error in planning gated RT. In conclusion, individualization of gating windows reduces the subjectivity in respiration-gated RT and improves the treatment of moving targets.Keywords
This publication has 27 references indexed in Scilit:
- Automated contour mapping using sparse volume sampling for 4D radiation therapyMedical Physics, 2007
- Four‐dimensional cone‐beam computed tomography using an on‐board imagerMedical Physics, 2006
- Deblurring of breathing motion artifacts in thoracic PET images by deconvolution methodsMedical Physics, 2006
- MO‐E‐330A‐07: Knowledge‐Based Auto‐Contouring in 4D Radiation TherapyMedical Physics, 2006
- Model‐based image reconstruction for four‐dimensional PETMedical Physics, 2006
- Dosimetric benefits of respiratory gating: a preliminary studyJournal of Applied Clinical Medical Physics, 2004
- A method for the reconstruction of four‐dimensional synchronized CT scans acquired during free breathingMedical Physics, 2003
- Respiration‐correlated spiral CT: A method of measuring respiratory‐induced anatomic motion for radiation treatment planningMedical Physics, 2002
- Improvement of CT-based treatment-planning models of abdominal targets using static exhale imagingInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 1998
- Respiration gated radiotherapy treatment: a technical studyPhysics in Medicine & Biology, 1996