Development of a novel high quantum efficiency MV x‐ray detector for image‐guided radiotherapy: A feasibility study
Authors:
Open Access
- 4 November 2019
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 47 (1) , 152-163
Abstract
Purpose To develop a new scintillating‐fiber‐based electronic portal imaging device (EPID) with a high quantum efficiency (QE) while preserving an adequate spatial resolution. Methods Two prototypes were built: one with a single pixel readout and the other with an active matrix flat panel imager (AMFPI) for readout. The energy conversion layer of both prototypes was made of scintillating‐fiber layers interleaved with corrugated lead sheets to form a honeycomb pattern. The scintillating fibers have a diameter of 1 mm and the distance between the centers of neighboring fibers on the same layer is 1.35 mm. The layers have 1.22 mm spacing between them. The energy conversion layer has a thickness of 2 cm. The modulation transfer function (MTF), anti‐scatter properties and sensitivity of the detector with a single pixel readout were measured using 6 MV beam on a LINAC machine. In addition, a Monte Carlo simulation was conducted to calculate the zero‐frequency detective quantum efficiency (DQE(0)) of the proposed detector with an active matrix flat panel imager for readout. Results The DQE(0) of the proposed detector can be 11.5%, which is about an order of magnitude higher than that of current EPIDs. The frequency of 50% modulation (f50) of the measured MTF is 0.2 mm‐1 at 6 MV, which is comparable to that of video‐based EPIDs. The scatter to primary ratio (SPR) measured with the detector at 10 cm air gap and 20×20 cm2 field size is approximately 30% lower than that of ionization‐chamber‐based detectors with a comparable QE. The detector noise which includes the x‐ray quantum noise and absorption noise is much larger than the electronic noise per pixel of the flat‐panel imager at a dose of less than two LINAC pulses. Thus the proposed detector is quantum noise limited down to very low doses (~ a couple of radiation pulses of the LINAC). A proof‐of‐concept image has been obtained using a 6 MV beam. Conclusions This work indicates that by using scintillating fibers and lead layers it is possible to increase the thickness of the detecting materials, and therefore the QE or the DQE(0) of the detector, while maintaining an adequate spatial resolution for MV x‐ray imaging. Due to the use of lead as the spacing material, the new detector also has anti‐scatter property, which will help improve the signal‐to‐noise ratio of the images. Further investigation to optimize the design of the detector and achieve a better combination of DQE and spatial resolution is warranted.Keywords
Funding Information
- Natural Sciences and Engineering Research Council of Canada (RGPIN‐311780, SAPJ‐326516)
This publication has 40 references indexed in Scilit:
- Finding an improved amorphous-silicon x-ray flat-panel detector configuration for the in-line geometryPhysics in Medicine & Biology, 2013
- An inherent anti-scatter detector for megavoltage x-ray imagingPhysics in Medicine & Biology, 2013
- Monte Carlo simulation of a novel water-equivalent electronic portal imaging device using plastic scintillating fibersMedical Physics, 2012
- Megavoltage X-Ray Imaging Based on Cerenkov Effect: A New Application of Optical Fibres to Radiation TherapyInternational Journal of Optics, 2011
- Low-dose megavoltage cone-beam CT imaging using thick, segmented scintillatorsPhysics in Medicine & Biology, 2011
- Flat Panel Detectors for Digital Radiography2010
- A Monte Carlo investigation of Swank noise for thick, segmented, crystalline scintillators for radiotherapy imagingMedical Physics, 2009
- Optimization of image quality and dose for Varian aS500 electronic portal imaging devices (EPIDs)Physics in Medicine & Biology, 2007
- Electronic portal imaging based on Cerenkov radiation: A new approach and its feasibilityMedical Physics, 2006
- Detective quantum efficiency of imaging systems with amplifying and scattering mechanismsJournal of the Optical Society of America A, 1987