Dosimetry quality control based on percent depth dose rate variation for checking beam quality in radiotherapy
- 1 July 2020
- journal article
- review article
- Published by VM Media Group sp. z o.o in Reports of Practical Oncology and Radiotherapy
- Vol. 25 (4) , 484-488
Abstract
Recently, the quality management inside a radiotherapy department has been crucial to treat cancer efficiently. Thus, many international bodies recommend multiple methods to check in periodically the dosimetry quality beyond the depth of 10 cm as the beam quality index. However, they evade checking out the beam dosimetry quality on both the build-up dose and the electronic equilibrium regions. The objective of this study is to cover the overall variation of the percent depth dose (PDD) by including all sub-regions in the procedure evaluation of the beam quality. In this work, we have studied and examined the dosimetry quality by considering the whole PDD variation. The PDD rate is therefore introduced to determine accurately the quality as an overall notion in external beam radiotherapy according to the field size and photon beam energy. We have presented the reasons and methods to introduce particles contamination, such as electrons and low photon energy in this new approach. The latter enables us to figure the dosimetry quality by extending the International Atomic Energy Agency (IAEA) procedure at any field size less than 25 x 25 cm(2) under the current conditions without being limited to 10 x 10 cm(2) on the exponential decay region. (C) 2020 Greater Poland Cancer Centre. Published by Elsevier B.V. All rights reserved.Keywords
This publication has 6 references indexed in Scilit:
- Study of Possibility to Reduce Flattening Filter Volume for Increasing Energetic Photons for High Radiotherapy EfficiencyMoscow University Physics Bulletin, 2017
- Relative Attenuation and Beam Softening Study with Flattening Filter Volume Reduction: Monte Carlo StudyMoscow University Physics Bulletin, 2017
- Dosimetry Investigation and Evaluation for Removing Flattening Filter Configuration of Linac: Monte Carlo StudyMoscow University Physics Bulletin, 2017
- Task Group 142 report: Quality assurance of medical acceleratorsa)Medical Physics, 2009
- Determination of the quality index (Q) for photon beams at arbitrary field sizesMedical Physics, 2009
- Comparison of IPSM 1990 photon dosimetry code of practice with IAEA TRS‐398 and AAPM TG‐51.Journal of Applied Clinical Medical Physics, 2009