Comparison of the Epson Expression 1680 flatbed and the Vidar VXR‐16 Dosimetry PRO™ film scanners for use in IMRT dosimetry using Gafchromic and radiographic film
Authors:
- Ellen WilcoxEWEllen Wilcox
- George DaskalovGDGeorge Daskalov
- Lucy NedialkovaLNLucy Nedialkova
- 12 December 2006
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 34 (1) , 41-48
Abstract
Intensity‐modulated radiotherapy (IMRT) treatment plan verification is often done using Kodak EDR2 film and a Vidar Dosimetry PRO™ film digitizer. However, since many hospitals are moving towards a filmless environment, access to a film processor may not be available. Therefore, we have investigated a newly available EBT film for IMRT dosimetry. Planar IMRT dose distributions are delivered to both EBT and EDR2 film and scanned with the Vidar VXR‐16 as well as an Epson Expression 1680 flatbed scanner. The measured dose distributions are then compared to those calculated with a Pinnacle treatment planning system. The IMRT treatments consisted of 7–9 beams for treatment of prostate, head and neck, and a few other sites. The films were analyzed using FilmQA™ (3cognition LLC) software. Comparisons between measured and calculated dose distributions are reported as dose difference (DD) (pixels within ), distance to agreement (DTA) , as well as gamma values (, ). Using EDR2 with the Vidar scanner is an established technique and agreement between calculated and measured dose distributions was better than 90% in all indices (DD, DTA, and ). However, agreement with calculations deteriorated reaching the lower 80% for EBT film scans with the Vidar scanner in logarithmic mode. The EBT Vidar scans obtained in linear mode showed an improved agreement to the upper 80% range, but artifacts were still observed across the scan. These artifacts were very distinct in all EBT scans and can be attributed to the way the film is transported through the scanner. In the Epson scanner both films are rigidly immobilized and the light source scans over the film. It was found that the Epson scanner performed equally well with both types of film giving agreement to better than 90% in all indices.Keywords
This publication has 44 references indexed in Scilit:
- Evaluation of a 2D diode array for IMRT quality assuranceRadiotherapy and Oncology, 2004
- A study of GafChromic XR Type R film response with reflective‐type densitometers and economical flatbed scannersJournal of Applied Clinical Medical Physics, 2003
- Dosimetric properties of an amorphous silicon electronic portal imaging device for verification of dynamic intensity modulated radiation therapyMedical Physics, 2003
- The use of a reflective scanner to study radiochromic film responsePhysics in Medicine & Biology, 2002
- Film dosimetry for intensity modulated radiation therapy: Dosimetric evaluationMedical Physics, 2002
- Validation of a precision radiochromic film dosimetry system for quantitative two-dimensional imaging of acute exposure dose distributionsMedical Physics, 2000
- A monitor unit verification calculation in intensity modulated radiotherapy as a dosimetry quality assuranceMedical Physics, 2000
- Radiochromic film dosimetry: Recommendations of AAPM Radiation Therapy Committee Task Group 55Medical Physics, 1998
- Quantitative evaluation of radiochromic film response for two‐dimensional dosimetryMedical Physics, 1997
- High-resolution dosimetry using radiochromic film and a document scannerPhysics in Medicine & Biology, 1996