A tandem calibration method for kerma–area product meters
Authors:
- P ToroiPTP Toroi
- T KomppaTKT Komppa
- A KosunenAKA Kosunen
- 18 August 2008
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 53 (18) , 4941-4958
Abstract
For measurements of the kerma-area product (KAP) in diagnostic x-ray imaging, a method for calibrating field KAP meters with a reference KAP meter is presented. In this tandem calibration method, the field KAP chamber is positioned similarly as in measurements with patients. The reference KAP chamber is placed at a specified distance and used in the x-ray beam simultaneously with the field KAP chamber. The tandem method provides a feasible and practical way for calibrating field KAP meters of any type in their clinical position. Accurate measurements of the irradiation geometry are not required, but comprehensive calibration for the reference KAP meter is needed.Keywords
This publication has 7 references indexed in Scilit:
- Monte Carlo study of the dependence of the KAP-meter calibration coefficient on beam aperture, x-ray tube voltage and reference planePhysics in Medicine & Biology, 2007
- COMPARISON OF TWO DOSE-AREA-PRODUCT IONIZATION CHAMBERS WITH DIFFERENT CONDUCTIVE SURFACE COATING FOR OVER-TABLE AND UNDER-TABLE TUBE CONFIGURATIONSHealth Physics, 2000
- Ionization chambers for measuring air kerma integrated over beam area. Deviations in calibration values using simplified calibration methodsPhysics in Medicine & Biology, 1998
- Transmission ionization chambers for measurements of air collision kerma integrated over beam area. Factors limiting the accuracy of calibrationPhysics in Medicine & Biology, 1996
- Recommendations on performance characteristics of diagnostic exposure meters: Report of AAPM Diagnostic X‐Ray Imaging Task Group No. 6Medical Physics, 1992
- An evaluation of the Diamentor transmission ionisation chamber in indicating exposure-area product (R cm2) during diagnostic radiological examinationsPhysics in Medicine & Biology, 1982
- Computation of bremsstrahlung X-ray spectra and comparison with spectra measured with a Ge(Li) detectorPhysics in Medicine & Biology, 1979