A cross‐sectional study of the radiation dose and image quality of X‐ray equipment used in IVR
Open Access
- 8 July 2016
- journal article
- research article
- Published by Wiley in Journal of Applied Clinical Medical Physics
- Vol. 17 (4) , 391-401
Abstract
There are case reports of injuries caused by the radiation from interventional radiology (IVR) X‐ray systems. Therefore, the management of radiation doses in IVR is important. However, no detailed report has evaluated image quality for a large number of IVR X‐ray systems. As a result, it is unclear whether the image quality of the X‐ray equipment currently used in IVR procedures is optimal. We compared the entrance surface doses and image quality of multiple IVR X‐ray systems. This study was conducted in 2014 at 13 medical facilities using 18 IVR X‐ray systems. We evaluated image quality and simultaneously measured the radiation dose. Entrance surface doses for fluoroscopy (duration, 1 min) and cineradiography (duration, 10 s) are measured using a 20‐cm‐thick acrylic plate and skin dose monitor. The image quality (such as spatial resolution and low‐contrast detectability) of both fluoroscopy and cineradiography was evaluated using a QC phantom. For fluoroscopy, the average entrance surface dose using the 20‐cm‐thick acrylic plate was 13.9 (range 2.1–28.2) mGy/min. For cineradiography, the average entrance surface dose was 24.6 (range 5.1–49.3) mGy/10 s. We found positive correlations between radiation doses and image quality scores, in general, especially for fluoroscopy. The differences in surface dose among the 18 IVR X‐ray systems were high (max/min, 9.7‐fold for cineradiography; 13.4‐fold for fluoroscopy). The differences in image quality scores (spatial resolution, low‐contrast detectability, and dynamic range) were also very large. In general, there tended to be a correlation between radiation dose and image quality. Periodical measurements of the radiation dose and image quality of the X‐ray equipment used for cineradiography and fluoroscopy in IVR are necessary. The need to minimize patient exposure requires that the dose be reduced to the minimum level that will generate an image with an acceptable degree of noise.PACS number(s): 87.57.C, 87.57.uq, 87.59.B, 87.59.bf, 87.59.C, 87.59.cf, 87.59.DjKeywords
Funding Information
- Grant-in-Aid for Scientific Research (24300179, 26460716)
This publication has 29 references indexed in Scilit:
- Quality Control Phantom for Flat Panel Detector X-ray SystemsHealth Physics, 2013
- The necessity of follow-up for radiation skin injuries in patients after percutaneous coronary interventions: radiation skin injuries will often be overlooked clinicallyActa Radiologica, 2012
- Comparison of the radiation dose in a cardiac IVR X-ray systemRadiation Protection Dosimetry, 2010
- Radiation Dose to the Pediatric Cardiac Catheterization and Intervention PatientAmerican Journal of Roentgenology, 2010
- Radiation Dose of Interventional Radiology System Using a Flat-Panel DetectorAmerican Journal of Roentgenology, 2009
- Radiation Exposure to Patients During Interventional Procedures in 20 Countries: Initial IAEA Project ResultsAmerican Journal of Roentgenology, 2009
- Preface, Executive Summary and GlossaryAnnals of the ICRP, 2007
- Relationship Between Fluoroscopic Time, Dose–Area Product, Body Weight, and Maximum Radiation Skin Dose in Cardiac Interventional ProceduresAmerican Journal of Roentgenology, 2006
- Does Digital Acquisition Reduce Patients' Skin Dose in Cardiac Interventional Procedures? An Experimental StudyAmerican Journal of Roentgenology, 2004
- Abstract: Avoidance of radiation injuries from medical interventional procedures, ICRP Publication 85Annals of the ICRP, 2000