Dosimetric characterization of a new surface‐conforming electron MLC prototype
Open Access
- 19 October 2023
- journal article
- research article
- Published by Wiley in Journal of Applied Clinical Medical Physics
- Vol. 25 (2) , e14173
Abstract
The purpose is to reduce normal tissue radiation toxicity for electron therapy through the creation of a surface‐conforming electron multileaf collimator (SCEM). The SCEM combines the benefits of skin collimation, electron conformal radiotherapy, and modulated electron radiotherapy.An early concept for the SCEM was constructed. It consists of leaves that protrude towards the patient, allowing the leaves to conform closely to irregular patient surfaces. The leaves are made of acrylic to decrease bremsstrahlung, thereby decreasing the out‐of‐field dose. Water tank scans were performed with the SCEM in place for various field sizes for all available electron energies (6, 9, 12, and 15 MeV) with a 0.5 cm air gap to the water surface at 100 cm source‐to‐surface distance (SSD). These measurements were compared with Cerrobend cutouts with the field size‐matched at 100 and 110 cm SSD. Output factor measurements were taken in solid water for each energy at dmax for both the cerrobend cutouts and SCEM at 100 cm SSD.Percent depth dose (PDD) curves for the SCEM shifted shallower for all energies and field sizes. The SCEM also produced a higher surface dose relative to Cerrobend cutouts, with the maximum being a 9.8% increase for the 3 cm × 9 cm field at 9 MeV. When compared to the Cerrobend cutouts at 110 cm SSD, the SCEM showed a significant decrease in the penumbra, particularly for lower energies (i.e., 6 and 9 MeV). The SCEM also showed reduced out‐of‐field dose and lower bremsstrahlung production than the Cerrobend cutouts. The SCEM provides significant improvement in the penumbra and out‐of‐field dose by allowing collimation close to the skin surface compared to Cerrobend cutouts. However, the added scatter from the SCEM increases shallow PDD values. Future work will focus on reducing this scatter while maintaining the penumbra and out‐of‐field benefits the SCEM has over conventional collimation.Keywords
This publication has 22 references indexed in Scilit:
- Characterization of an extendable multi-leaf collimator for clinical electron beamsPhysics in Medicine & Biology, 2011
- Comparison of electron IMRT to helical photon IMRT and conventional photon irradiation for treatment of breast and chest wall tumoursRadiotherapy and Oncology, 2010
- Delivery of modulated electron beams with conventional photon multi-leaf collimatorsPhysics in Medicine & Biology, 2008
- Dose to the Contralateral Breast From Radiotherapy and Risk of Second Primary Breast Cancer in the WECARE StudyInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2008
- Characterization of an add-on multileaf collimator for electron beam therapyPhysics in Medicine & Biology, 2008
- Review of electron beam therapy physicsPhysics in Medicine & Biology, 2006
- Characterization of megavoltage electron beams delivered through a photon multi-leaf collimator (pMLC)Physics in Medicine & Biology, 2006
- The role of radiotherapy in cancer treatmentCancer, 2005
- Utilization of custom electron bolus in head and neck radiotherapyJournal of Applied Clinical Medical Physics, 2003
- Computer-aided design and fabrication of an electron bolus for treatment of the paraspinal musclesInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 1995