Computed tomography employing sensing of high energy particle current
- 17 October 2024
- journal article
- research article
- Published by IOP Publishing in Biomedical Physics & Engineering Express
- Vol. 10 (6) , 065042
Abstract
Objective. We demonstrate High Energy Current Computed Tomography (HEC-CT) employing megavoltage linac x-rays. Approach. Using deterministic radiation transport we simulate two-parameter HEC-CT projections and using inverse Fourier transform we reconstruct two distinct material parameters for water phantoms with ICRP tissue inserts and materials of different atomic number Z. The HEC-CT projections are obtained by beam scanning and rotating the object. Main Results. The first HEC-CT material parameter is alike the standard attenuation coefficient with dependence on atomic number and material density similar to the Hounsfield Units. The second material parameter has opposite trends and does not find any analogy in the standard CT framework. Significance. New CT method has been invented for medical imaging or non-destructive testing. The key feature of the technique is a two-value CT reconstruction based on particle current instead of transmitted dose.Keywords
This publication has 11 references indexed in Scilit:
- Remote sensing of high energy charged particle current (HEC) for megavoltage therapeutic electron beamsPhysics in Medicine & Biology, 2024
- Remote sensing of high energy particle currents (HEC): towards HEC radiographyJournal of Physics D: Applied Physics, 2024
- Dual- and multi-energy CT for particle stopping-power estimation: current state, challenges and potentialPhysics in Medicine & Biology, 2023
- Self‐powered nano‐porous aerogel x‐ray sensor employing fast electron currentMedical Physics, 2019
- Towards thin-film self-powered radiation detectors employing disparate conductive layersJournal of Physics D: Applied Physics, 2015
- PrefaceJournal of the ICRU, 2011
- Megavoltage cone beam computed tomography: Commissioning and evaluation of patient doseJournal of Medical Physics, 2011
- Calibration of megavoltage cone‐beam CT for radiotherapy dose calculations: Correction of cupping artifacts and conversion of CT numbers to electron densityMedical Physics, 2008
- Megavoltage cone-beam CT: System description and clinical applicationsMedical Dosimetry, 2006
- The measurement of proton stopping power using proton-cone-beam computed tomographyPhysics in Medicine & Biology, 2000