Dual‐echo Z‐shimmed proton resonance frequency‐shift magnetic resonance thermometry near metallic ablation probes: Technique and temperature precision
- 10 February 2017
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 78 (6) , 2299-2306
Abstract
Purpose: To improve the precision of proton resonance frequency‐shift magnetic resonance thermometry near ablation probes by recovering near‐probe image signals that are typically lost due to magnetic susceptibility‐induced field distortions.Methods: A dual‐echo gradient‐recalled echo sequence was implemented, in which the first echo was under‐ or over‐refocused in the slice dimension to recover image signal and temperature precision near a probe, and the second echo was fully refocused to obtain image signal everywhere else in the slice. A penalized maximum likelihood algorithm was implemented to estimate a single temperature map from both echoes. Agar phantom and ex vivo experiments with and without microwave heating at 3 T evaluated how much temperature precision was improved near a microwave ablator compared to a conventional single‐echo scan as a function of slice and needle orientation in the magnet.Results: The number of near‐probe voxels with temperature standard deviation C was decreased by 51% in the phantom experiment, averaged across orientations, and by 31% in the pork. Temperature maps near the probe were more smoother and more complete in all orientations.Conclusion: Dual‐echo z‐shimmed temperature imaging can recover image signal for more precise temperature mapping near metallic ablation probes. Magn Reson Med 78:2299–2306, 2017. © 2017 International Society for Magnetic Resonance in Medicine.Keywords
Funding Information
- National Institutes of Health (R21 NS091735, T32 EB021937)
This publication has 20 references indexed in Scilit:
- Heating induced near deep brain stimulation lead electrodes during magnetic resonance imaging with a 3 T transceive volume head coilPhysics in Medicine & Biology, 2012
- Center‐out radial sampling with off‐resonant reconstruction for efficient and accurate localization of punctate and elongated paramagnetic structuresMagnetic Resonance in Medicine, 2012
- Metal-Induced Artifacts in MRIAmerican Journal of Roentgenology, 2011
- Magnetic resonance imaging near metal implantsJournal of Magnetic Resonance Imaging, 2010
- MR thermometryJournal of Magnetic Resonance Imaging, 2008
- Reduction of blurring in view angle tilting MRIMagnetic Resonance in Medicine, 2005
- Passive tracking exploiting local signal conservation: The white marker phenomenonMagnetic Resonance in Medicine, 2003
- Functional MR imaging using gradient‐echo echo‐planar imaging in the presence of large static field inhomogeneitiesJournal of Magnetic Resonance Imaging, 1995
- Reduction of susceptibility artifact in gradient‐echo imagingMagnetic Resonance in Medicine, 1992
- Direct FLASH MR imaging of magnetic field inhomogeneities by gradient compensationMagnetic Resonance in Medicine, 1988