Initial testing of pegfilgrastim (Neulasta Onpro) on‐body injector in multiple radiological imaging environments
Authors:
Open Access
- 4 January 2021
- journal article
- research article
- Published by Wiley in Journal of Applied Clinical Medical Physics
- Vol. 22 (1) , 343-349
Abstract
Purpose: An increasing number of implantable or external devices can impact whether patients can receive radiological imaging examinations. This study examines and tests the Neulasta (pegfilgrastim) Onpro on‐body injector in multiple imaging environments.Methods: The injector was analyzed for four imaging modalities with testing protocols and strategies developed for each modality. In x‐ray and computed tomography (CT), scans with much higher exposure than clinical protocols were performed with the device attached to an anthropomorphic phantom. The device was monitored until the completion of drug delivery. For magnetic resonance imaging (MRI), the device was assessed using a hand‐held magnet and underwent the magnetically induced displacement testing in a 1.5T clinical MRI scanner room. For ultrasound, magnetic field changes were measured around an ultrasound scanner system with three transducers.Results: For x‐ray and CT no sign of device error was identified during or after the high radiation exposure scans. Drug delivery was completed at expected timing with expected volume. For MRI the device showed significant attractive force towards the hand‐held magnet and a 50‐degree deflection angle at 50 cm from the opening of the scanner bore. No further assessment from the gradient or radiofrequency field was deemed necessary. For ultrasound the maximum magnetic field change from baseline was measured to be +11.7 μT in comparison to +74.2 μT at 4 inches from a working microwave.Conclusions: No device performance issue was identified under the extreme test conditions in x‐ray or CT. The device was found to be MR Unsafe. Magnetic field changes around an ultrasound system met the limitation set by manufacture. Patient ultrasound scanning is considered safe as long as the transducers do not inadvertently loosen the device.Keywords
This publication has 25 references indexed in Scilit:
- MRI in Patients with Cardiac Implantable Electronic DevicesRadiology, 2018
- 2017 HRS expert consensus statement on magnetic resonance imaging and radiation exposure in patients with cardiovascular implantable electronic devicesHeart Rhythm, 2017
- Magnetic resonance imaging safety in pacemaker and implantable cardioverter defibrillator patients: how far have we come?European Heart Journal, 2015
- Safety of Computed Tomography in Patients With Cardiac Rhythm Management DevicesJACC, 2014
- ACR guidance document on MR safe practices: 2013Journal of Magnetic Resonance Imaging, 2013
- Computed Tomography Scan and ICD InteractionCase Reports in Medicine, 2009
- Effect of Diagnostic X-rays on Implantable Cardiac Pacemakers and Implantable Cardioverter Defibrillators, and Its ManagementJapanese Journal of Radiological Technology, 2008
- Safety of Magnetic Resonance Imaging in Patients With Cardiovascular DevicesCirculation, 2007
- Effects of CT Irradiation on Implantable Cardiac Rhythm Management Devices1Radiology, 2007
- Does High-Power Computed Tomography Scanning Equipment Affect the Operation of Pacemakers?Circulation Journal, 2006