Abstract
Lorentz force electrical impedance tomography (LFEIT) combines ultrasound stimulation and electromagnetic field detection with the goal of creating a high contrast and high resolution hybrid imaging modality. To reduce the peak stimulation power to the ultrasound transducer in LFEIT, linearly frequency-modulated ultrasound pulse was investigated in this work. First, the coherency between linearly frequency-modulated ultrasound excitation and the resulting local current density was established theoretically. Then, experiments were done using different agar phantoms of conductivity ranging from 0.2 S/m to 0.5 S/m. The results showed: (1) using electrical signal of peak instantaneous power of 39.54 dBm to the ultrasound transducer, which was 25.5 dB lower than the peak instantaneous power of the high-voltage narrow pulse adopted in traditional LFEIT (65.05 dBm), the linearly frequency-modulated ultrasound pulse based LFEIT can detect the electrical conductivity discontinuity positions precisely; (2) the reconstructed B-scan image of the electrical conductivity discontinuity distribution is comparable to that obtained through LFEIT with high-voltage narrow pulse; (3) axial resolution of 1 mm was achieved with the experimental setup. The method of linearly frequency-modulated ultrasound pulse stimulation and coherent detection initiates an alternative scheme towards the clinical application of LFEIT.