Results of a phase I, non-randomized study evaluating a Magnetic Occult Lesion Localization Instrument (MOLLI) for excision of non-palpable breast lesions
Authors:
Open Access
- 21 November 2019
- journal article
- clinical trial
- Published by Springer Nature in Breast Cancer Research and Treatment
- Vol. 179 (3) , 671-676
Abstract
Purpose: Magnetic Occult Lesion Localization Instrument (MOLLI) is a wireless, non-radioactive alternative for non-palpable breast lesion localization. The primary objective of this first-in-human study was to evaluate the clinical feasibility of using MOLLI for intraoperative localization of non-palpable breast lesions. Methods: Twenty women with non-palpable breast lesions at a single institution received a lumpectomy using the MOLLI guidance system. Patients were co-localized with magnetic and radioactive markers up to 7 days before excision by a dedicated breast radiologist under sonographic guidance. Both markers were localized intraoperatively using dedicated hand-held probes. The primary outcome was successful excision of the magnetic marker, confirmed radiographically and pathologically. Demographic data, margin positivity, and re-excision rates were collected. Surgical oncologists, radiologists, and pathology staff were surveyed for user satisfaction. Results: Post-radiological analysis: Post-implant mammograms verified that 17/20 markers were placed directly in the lesion center. Radiologists reported that all marker implantations procedures were “easy” or “very easy” following a single training session. Post-surgical analysis: All MOLLI markers were successfully removed with the specimen during surgical excision. In all cases, surgeons ranked the MOLLI guidance system as “very easy” for lesion localization. Pathologic analysis: All patients had negative margins. All anatomic pathology staff ranked the MOLLI system as “very easy” to localize markers. Conclusions: The MOLLI guidance system is a reliable and accurate method for intraoperative localization of non-palpable breast lesions. Further evaluation of the MOLLI system in studies against current standards of care is required to demonstrate system cost-effectiveness and improved patient-reported outcomes.Keywords
Funding Information
- Canadian Cancer Society Research Institute (705239)
- Sunnybrook Research Institute
This publication has 14 references indexed in Scilit:
- Beyond Wires and Seeds: Reflector-guided Breast Lesion Localization and ExcisionRadiology, 2017
- Update on Preoperative Breast LocalizationRadiologic Clinics of North America, 2017
- Long-term health status as measured by EQ-5D among patients with metastatic breast cancer: comparison of first-line oral S-1 and taxane therapies in the randomized phase III SELECT BC trialQuality of Life Research, 2016
- Radioactive Seed Localization Versus Wire Localization for Lumpectomies: A Comparison of OutcomesAmerican Journal of Roentgenology, 2015
- Radioactive Seed Localization for Nonpalpable Breast Lesions: Review of 1,000 Consecutive Procedures at a Single InstitutionAnnals of Surgical Oncology, 2011
- Radioactive Seed Localization of Breast Lesions: An Adequate Localization Method without Seed MigrationThe Breast Journal, 2011
- Early breast cancerThe Lancet, 2009
- Radioactive Seed Localization Breast Biopsy and Lumpectomy: Can Specimen Radiographs Be Eliminated?Annals of Surgical Oncology, 2003
- EuroQol: the current state of playHealth Policy, 1996
- Complications of Needle Localization of Foreign Bodies and Nonpalpable Breast LesionsArchives of Surgery, 1988