Evaluating Image Quality and Radiation Dose in Low-Dose Thoraco-Abdominal CT Angiography with a Tin Filter for Patients with Aortic Disease
Open Access
- 8 February 2024
- journal article
- research article
- Published by MDPI AG in Journal of Clinical Medicine
- Vol. 13 (4) , 996
Abstract
Background: We aimed to compared radiation exposure and image quality between tin-filter-based and standard dose thoraco-abdominal computed tomography angiography (TACTA) protocols, aiming to address a gap in the existing literature. Methods: In this retrospective study, ninety consecutive patients undergoing TACTA were included. Of these, 45 followed a routine standard-dose protocol (ST100kV), and 45 underwent a low-dose protocol with a tin filter (TF100kV). Radiation metrics were compared. The signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR), and figure of merit (FOM) were calculated for the thoracic and abdominal aorta and right common iliac artery. Two independent readers assessed the image noise, image contrast, sharpness, and subjective image quality. Results: The mean dose for the TF100kV group was significantly lower (DLP 128.25 ± 18.18 mGy*cm vs. 662.75 ± 181.29, p < 0.001; CTDIvol 1.83 ± 0.25 mGy vs. 9.28 ± 2.17, p = 0.001), with an effective dose close to 2.3 mSv (2.31 ± 0.33 mSv; p < 0.001). The TF100kV group demonstrated greater dose efficiency (FOM, thoracic aorta: 36.70 ± 22.77 vs. 13.96 ± 13.18 mSv−1, p < 0.001) compared to the ST100kV group. Conclusions: Dedicated low-dose TACTA using a tin filter can significantly reduce the radiation dose while maintaining sufficient diagnostic image quality.Keywords
Funding Information
- Ewha Womans University Research Grant
This publication has 18 references indexed in Scilit:
- Effect of tin filter-based spectral shaping CT on image quality and radiation dose for routine use on ultralow-dose CT protocols: A phantom studyDiagnostic and Interventional Imaging, 2020
- CT iterative reconstruction algorithms: a task-based image quality assessmentEuropean Radiology, 2019
- Common artefacts encountered on images acquired with combined compressed sensing and SENSEInsights into Imaging, 2018
- The evolution of image reconstruction for CT—from filtered back projection to artificial intelligenceEuropean Radiology, 2018
- Prospective evaluation of ultra-low-dose contrast-enhanced 100-kV abdominal computed tomography with tin filter: effect on radiation dose reduction and image quality with a third-generation dual-source CT systemEuropean Radiology, 2018
- High-pitch low-voltage CT coronary artery calcium scoring with tin filtration: accuracy and radiation dose reductionEuropean Radiology, 2018
- CT coronary calcium scoring with tin filtration using iterative beam-hardening calcium correction reconstructionEuropean Journal of Radiology, 2017
- Unenhanced third-generation dual-source chest CT using a tin filter for spectral shaping at 100 kVpEuropean Journal of Radiology, 2015
- Chest CT using spectral filtration: radiation dose, image quality, and spectrum of clinical utilityEuropean Radiology, 2014
- Accuracy and precision of vessel area assessment: Manual versus automatic lumen delineation based on full‐width at half‐maximumJournal of Magnetic Resonance Imaging, 2012