We study the optimum operating conditions for a rotating retarder fixed polarizer (RRFP) when the measurements are not quasi-instantaneous but time-averaged. We obtain the optimum retardance and retarder orientations as a function of the integrated angle interval. We also study how the increase in the number of time-averaged measurements leads to a better equally weighted variance (EWV) value, and thus, to a better performance of the polarimeter in terms of noise amplification for the case of additive noise. Two different analyzers configurations are studied in this work: uniformly spaced retarder angles and when measurements are taken at optimum angles (non-uniformly spaced angles). We also consider the case of polychromatic illumination. We discuss the best measurement conditions in terms of the signal-to-noise ratio depending on whether there is a fixed or a limited amount of photons per measurement.