FEA Modeling of Soft Tissue Interaction for Active Needles With a Rotational Tip Joint
Authors:
Open Access
- 21 April 2022
- journal article
- research article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Access
- Vol. 10, 46291-46301
Abstract
A finite element analysis (FEA) model was developed for investigating the design and the tissue interaction of actively steerable needles with a rotational tip joint in soft tissue, based on the coupled Eulerian-Lagrangian (CEL) mesh. The new model is algorithmically simple in that the proposed CEL strategy allows needle insertion simulations along non-predetermined paths in soft tissue without the re-mesh at every direction change of the needle tip. For the features, various steering motions can be simulated and studied for different needle designs in a simpler approach by changing the needle geometry and boundary conditions. The developed FEA model, using the thoroughly measured material properties, was validated for predicting insertion path and estimating the tissue interaction forces inside two different gelatin tissue phantoms. Further, using the validated model, the effect of tip geometry on tissue was briefly investigated. Given bevel angles, it was found that the ratio of tip length to the diameter dominates the tissue damage gradient. The results demonstrate that the proposed model effectively examines the steering and tissue insertion of actively steerable needles and investigates the tip design to minimize the tissue damage by the needle steering.Keywords
Funding Information
- National Research Foundation of Korea (NRF) grant funded by the Korean government (2021R1A2C2011854)
- Brain Korea 21 Four Program
- Ewha Womans University Research (2019)
- Ewha Womans University (2021)
This publication has 26 references indexed in Scilit:
- Controlling the error on target motion through real‐time mesh adaptation: Applications to deep brain stimulationInternational Journal for Numerical Methods in Biomedical Engineering, 2018
- Design and Fabrication of Laser-Machined Hinge Joints on Miniature Tubes for Steerable Medical DevicesJournal of Mechanisms and Robotics, 2017
- Robotic-Assisted Needle Steering Around Anatomical Obstacles Using Notched Steerable NeedlesIEEE Journal of Biomedical and Health Informatics, 2017
- Real-Time Error Control for Surgical SimulationIEEE Transactions on Biomedical Engineering, 2017
- Simulation and experimental studies in needle–tissue interactionsJournal of Clinical Monitoring and Computing, 2016
- Dynamic analysis of a needle insertion for soft materials: Arbitrary Lagrangian–Eulerian-based three-dimensional finite element analysisComputers in Biology and Medicine, 2014
- Cutting Force of Hollow Needle Insertion in Soft Tissue2013
- Detailed finite element modelling of deep needle insertions into a soft tissue phantom using a cohesive approachComputer Methods in Biomechanics and Biomedical Engineering, 2013
- Application of a Coupled Eulerian–Lagrangian approach on geomechanical problems involving large deformationsComputers and Geotechnics, 2011
- Simulations of Needle Insertion by Using a Eulerian Hydrocode FEM and the Experimental Validations2008