Effect of 7.0 Tesla MRI on Upper Eyelid Implants
- 1 November 2006
- journal article
- other
- Published by Wolters Kluwer Health in Ophthalmic Plastic & Reconstructive Surgery
- Vol. 22 (6) , 480-482
Abstract
To ensure the MRI compatibility of various eyelid implants in high-field MRI, 3 eyelid weights made of pure gold (99.99%), pure platinum (99.95%), and a platinum (97%)/iridium (3%) alloy were examined in vitro. Temperature changes, position changes, and imaging artifacts of the different implants were determined in a small-bore 7.0 Tesla MRI system. The 7.0 Tesla MRI system demonstrated that none of the eyelid implants carried a risk of heating or dislocation; therefore, these implants are MRI compatible up to a magnetic field strength of 7.0 Tesla.Keywords
This publication has 5 references indexed in Scilit:
- MR-KoronarangiographieRöFo - Fortschritte auf dem Gebiet der Röntgenstrahlen und der bildgebenden Verfahren, 2004
- Effect of Magnetic Resonance Imaging on Implantable Eyelid WeightsAnnals of Otology, Rhinology & Laryngology, 1995
- Gold weight or spring implants and MRIThe Laryngoscope, 1993
- Heating effects of metallic implants by MRI examinationsMagnetic Resonance in Medicine, 1988