Long-term Inter-protocol kV CBCT image quality assessment for a ring-gantry linac via automated QA approach
- 1 January 2020
- journal article
- research article
- Published by IOP Publishing in Biomedical Physics & Engineering Express
- Vol. 6 (1) , 015025
Abstract
We develop a fully automated QA process to compare the image quality of all kV CBCT protocols on a Halcyon linac with ring gantry design, and evaluate image quality stability over a 10-month period. A total of 19 imaging scan and reconstruction protocols were characterized with measurement on a newly released QUART phantom. A set of image analysis algorithms were developed and integrated into an automated analysis suite to derive key image quality metrics, including HU value accuracy on density inserts, HU uniformity using the background plate, high contrast resolution with the modulation transfer function (MTF) from the edge profiles, low contrast resolution using the signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR), slice thickness with the air gap modules, and geometric accuracy with the diameter of the phantom. Image quality data over 10 months was tracked and analyzed to evaluate the stability of the Halcyon kV imaging system. The HU accuracy over all 19 protocols is within tolerance (±50HU). The maximum uniformity deviation is 12.2 HU. The SNR and CNR, depending on the protocol selected, range from 18.5–911.9 and 1.9–102.8, respectively. A much-improved SNR and CNR were observed for iterative reconstruction (iCBCT) modes and protocols designed for large subjects over low dose and fast scanning modes. The Head and Image Gently protocols have the greatest high contrast resolution with MTF 10% over 1 lp/mm and MTF 50% over 0.6 lp/mm. The iCBCT mode slightly improved the MTF 10% and MTF 50% compared to the Feldkamp-Davis-Kress approach. The slice thickness (maximum error of 0.31 mm) and geometry metrics (maximum error of 0.7 mm) are all within tolerance (±0.5 mm for slice thickness and ±1 mm for geometry metrics). The long-term study over 10-month showed no significant drift for all key image quality metrics, which indicated the kV CBCT image quality is stable over time.Keywords
Funding Information
- Varian Medical Systems
This publication has 10 references indexed in Scilit:
- Characterization of a prototype rapid kilovoltage x‐ray image guidance system designed for a ring shape radiation therapy unitMedical Physics, 2019
- Real-time volumetric relative dosimetry for magnetic resonance—image-guided radiation therapy (MR-IGRT)Physics in Medicine & Biology, 2018
- WE‐AB‐207A‐08: BEST IN PHYSICS (IMAGING): Advanced Scatter Correction and Iterative Reconstruction for Improved Cone‐Beam CT Imaging On the TrueBeam Radiotherapy MachineMedical Physics, 2016
- A quality assurance framework for the fully automated and objective evaluation of image quality in cone‐beam computed tomographyMedical Physics, 2014
- Image Guided Radiation Therapy (IGRT) Technologies for Radiation Therapy Localization and DeliveryInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2013
- Quality assurance for image‐guided radiation therapy utilizing CT‐based technologies: A report of the AAPM TG‐179Medical Physics, 2012
- Task Group 142 report: Quality assurance of medical acceleratorsa)Medical Physics, 2009
- Clinical Experiences With Onboard Imager KV Images for Linear Accelerator–Based Stereotactic Radiosurgery and Radiotherapy SetupInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2009
- A quality assurance program for image quality of cone‐beam CT guidance in radiation therapyMedical Physics, 2008
- Clinical use of electronic portal imaging: Report of AAPM Radiation Therapy Committee Task Group 58Medical Physics, 2001