Design and implementation of quality control plans that integrate moving average and internal quality control: incorporating the best of both worlds
- 23 March 2019
- journal article
- research article
- Published by Walter de Gruyter GmbH in Clinical Chemistry and Laboratory Medicine (CCLM)
- Vol. 57 (9) , 1329-1338
Abstract
Background: New moving average quality control (MA QC) optimization methods have been developed and are available for laboratories. Having these methods will require a strategy to integrate MA QC and routine internal QC. Methods: MA QC was considered only when the performance of the internal QC was limited. A flowchart was applied to determine, per test, whether MA QC should be considered. Next, MA QC was examined using the MA Generator (www.huvaros.com), and optimized MA QC procedures and corresponding MA validation charts were obtained. When a relevant systematic error was detectable within an average daily run, the MA QC was added to the QC plan. For further implementation of MA QC for continuous QC, MA QC management software was configured based on earlier proposed requirements. Also, protocols for the MA QC alarm work-up were designed to allow the detection of temporary assay failure based on previously described experiences. Results: Based on the flowchart, 10 chemistry, two immunochemistry and six hematological tests were considered for MA QC. After obtaining optimal MA QC settings and the corresponding MA validation charts, the MA QC of albumin, bicarbonate, calcium, chloride, creatinine, glucose, magnesium, potassium, sodium, total protein, hematocrit, hemoglobin, MCH, MCHC, MCV and platelets were added to the QC plans. Conclusions: The presented method allows the design and implementation of QC plans integrating MA QC for continuous QC when internal QC has limited performance.Keywords
This publication has 17 references indexed in Scilit:
- Optimization of a Moving Averages Program Using a Simulated Annealing Algorithm: The Goal is to Monitor the Process Not the PatientsClinical Chemistry, 2016
- Optimization and validation of moving average quality control procedures using bias detection curves and moving average validation chartsClinical Chemistry and Laboratory Medicine (CCLM), 2016
- Selecting Statistical Procedures for Quality Control Planning Based on Risk ManagementClinical Chemistry, 2016
- A method for optimization and validation of moving average as continuous analytical quality control instrument demonstrated for creatinineClinica Chimica Acta, 2016
- Changing the paradigm of laboratory quality control through implementation of real-time test results monitoring: For patients by patientsClinical Biochemistry, 2015
- Practical application of Sigma Metrics QC procedures in clinical chemistryClinical Chemistry and Laboratory Medicine (CCLM), 2011
- Assessing the Impact of the Frequency of Quality Control Testing on the Quality of Reported Patient ResultsClinical Chemistry, 2008
- Laboratory Quality Control: Using Patient Data to Assess Analytical PerformanceClinical Chemistry and Laboratory Medicine (CCLM), 2003
- Performance Evaluation and Planning for Patient-Based Quality Control ProceduresAmerican Journal of Clinical Pathology, 2000
- Assessment of “Average of Normals” Quality Control Procedures and Guidelines for ImplementationAmerican Journal of Clinical Pathology, 1984