Abstract
This study investigates the effect of the pilot hole size, implant depth, synthetic bone density, and screw size on the pullout strength of the self-tapping screw using analytical, finite element, and experimental methodologies. Stress distribution and failure propagation mode around the implant thread zone are also investigated. Based on the finite element analysis (FEA) results, an analytical model for the pullout strength of the self-tapping screw is constructed in terms of the (synthetic) bone mechanical properties, screw size, and the implant depth. The pullout performance of self-tapping screws is discussed. Results from the analytical and finite element models are experimentally validated.Keywords
This publication has 17 references indexed in Scilit:
- Assessment of different screw augmentation techniques and screw designs in osteoporotic spinesEuropean Spine Journal, 2008
- Increasing Bending Strength and Pullout Strength in Conical Pedicle Screws: Biomechanical Tests and Finite Element AnalysesJournal of Spinal Disorders & Techniques, 2008
- Cortical Screw Pullout Strength and Effective Shear Stress in Synthetic Third Generation Composite FemursJournal of Biomechanical Engineering, 2006
- Increase of pullout strength of spinal pedicle screws with conical core: Biomechanical tests and finite element analysesJournal of Orthopaedic Research, 2005
- Effects of hole preparation on screw pullout resistance and insertional torque: a biomechanical studyJournal of Neurosurgery: Spine, 2001
- Factors Affecting the Pullout Strength of Cancellous Bone ScrewsJournal of Biomechanical Engineering, 1996
- Extraction Strength of Cannulated Cancellous Bone ScrewsJournal of Orthopaedic Trauma, 1993
- Comparison of bone screw holding strength in healthy bovine and osteoporotic human cancellous boneClinical Materials, 1992
- Influence of Bone Mineral Density on the Fixation of Thoracolumbar Implants A Comparative Study of Transpedicular Screws, Laminar Hooks, and Spinous Process WiresSpine, 1990
- Pedicle Screws: Axial Pull-out Strength in the Lumbar SpineSpine, 1988