Two‐dimensional ionization chamber arrays for IMRT plan verification

Authors:
  • ABArne Blechschmidt
    Arne Blechschmidt
  • ADArmand Djouguela
    Armand Djouguela
  • RKRalf Kollhoff
    Ralf Kollhoff
  • ARAntje Rubach
    Antje Rubach
  • KWKay C. Willborn
    Kay C. Willborn
  • DHDietrich Harder
    Dietrich Harder

Abstract
In this paper we describe a concept for dosimetric treatment plan verification using two‐dimensional ionization chamber arrays. Two different versions of the 2D‐ARRAY (PTW‐Freiburg, Germany) will be presented, a matrix of chambers (chamber cross section ; the distance between chamber centers, ) and a matrix of chambers (chamber cross section ; the distance between chamber centers is ). The two‐dimensional response function of a single chamber is experimentally determined by scanning it with a slit beam. For dosimetric plan verification, the expected two‐dimensional distribution of the array signals is calculated via convolution of the planned dose distribution, obtained from the treatment planning system, with the two‐dimensional response function of a single chamber. By comparing the measured two‐dimensional distribution of the array signals with the expected one, a distribution of deviations is obtained that can be subjected to verification criteria, such as the gamma index criterion. As an example, this verification method is discussed for one sequence of an IMRT plan. The error detection capability is demonstrated in a case study. Both versions of two‐dimensional ionization chamber arrays, together with the developed treatment plan verification strategy, have been found to provide a suitable and easy‐to‐handle quality assurance instrument for IMRT.