Two‐dimensional ionization chamber arrays for IMRT plan verification
Authors:
- Bjoern PoppeBPBjoern Poppe
- Arne BlechschmidtABArne Blechschmidt
- Armand DjouguelaADArmand Djouguela
- Ralf KollhoffRKRalf Kollhoff
- Antje RubachARAntje Rubach
- Kay C. WillbornKWKay C. Willborn
- Dietrich HarderDHDietrich Harder
- 22 March 2006
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 33 (4) , 1005-1015
Abstract
In this paper we describe a concept for dosimetric treatment plan verification using two‐dimensional ionization chamber arrays. Two different versions of the 2D‐ARRAY (PTW‐Freiburg, Germany) will be presented, a matrix of chambers (chamber cross section ; the distance between chamber centers, ) and a matrix of chambers (chamber cross section ; the distance between chamber centers is ). The two‐dimensional response function of a single chamber is experimentally determined by scanning it with a slit beam. For dosimetric plan verification, the expected two‐dimensional distribution of the array signals is calculated via convolution of the planned dose distribution, obtained from the treatment planning system, with the two‐dimensional response function of a single chamber. By comparing the measured two‐dimensional distribution of the array signals with the expected one, a distribution of deviations is obtained that can be subjected to verification criteria, such as the gamma index criterion. As an example, this verification method is discussed for one sequence of an IMRT plan. The error detection capability is demonstrated in a case study. Both versions of two‐dimensional ionization chamber arrays, together with the developed treatment plan verification strategy, have been found to provide a suitable and easy‐to‐handle quality assurance instrument for IMRT.Keywords
This publication has 24 references indexed in Scilit:
- Guidance document on delivery, treatment planning, and clinical implementation of IMRT: Report of the IMRT subcommittee of the AAPM radiation therapy committeeMedical Physics, 2003
- Evaluation of Kodak EDR2 film for dose verification of intensity modulated radiation therapy delivered by a static multileaf collimatorMedical Physics, 2002
- The value of radiographic film for the characterization of intensity-modulated beamsPhysics in Medicine & Biology, 2002
- Film dosimetry for intensity modulated radiation therapy: Dosimetric evaluationMedical Physics, 2002
- Gel dosimetry for the dose verification of Intensity Modulated Radiotherapy TreatmentsZeitschrift für Medizinische Physik, 2002
- Dosimetrische Verifikation von IMRT-Gesamtplänen am Deutschen Krebsforschungszentrum HeidelbergZeitschrift für Medizinische Physik, 2002
- Delivery of intensity‐modulated radiation therapy with a conventional multileaf collimator: Comparison of dynamic and segmental methodsMedical Physics, 2001
- Intensity-Modulated Radiation Therapy2001
- Inverse bestrahlungsplanungStrahlentherapie und Onkologie, 1999
- Realization and verification of three-dimensional conformal radiotherapy with modulated fieldsInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 1994