Time dependent pre-treatment EPID dosimetry for standard and FFF VMAT
Authors:
- Sebastiaan M J J G NijstenSNSebastiaan M J J G Nijsten
- Lucas C G G PersoonLPLucas C G G Persoon
- Stefan G ScheibSSStefan G Scheib
- Christof BaltesCBChristof Baltes
- Frank VerhaegenFVFrank Verhaegen
- 4 August 2014
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 59 (16) , 4749-4768
Abstract
Methods to calibrate Megavoltage electronic portal imaging devices (EPIDs) for dosimetry have been previously documented for dynamic treatments such as intensity modulated radiotherapy (IMRT) using flattened beams and typically using integrated fields. While these methods verify the accumulated field shape and dose, the dose rate and differential fields remain unverified. The aim of this work is to provide an accurate calibration model for time dependent pre-treatment dose verification using amorphous silicon (a-Si) EPIDs in volumetric modulated arc therapy (VMAT) for both flattened and flattening filter free (FFF) beams. A general calibration model was created using a Varian TrueBeam accelerator, equipped with an aS1000 EPID, for each photon spectrum 6 MV, 10 MV, 6 MV-FFF, 10 MV-FFF. As planned VMAT treatments use control points (CPs) for optimization, measured images are separated into corresponding time intervals for direct comparison with predictions. The accuracy of the calibration model was determined for a range of treatment conditions. Measured and predicted CP dose images were compared using a time dependent gamma evaluation using criteria (3%, 3 mm, 0.5 sec). Time dependent pre-treatment dose verification is possible without an additional measurement device or phantom, using the on-board EPID. Sufficient data is present in trajectory log files and EPID frame headers to reliably synchronize and resample portal images. For the VMAT plans tested, significantly more deviation is observed when analysed in a time dependent manner for FFF and non-FFF plans than when analysed using only the integrated field. We show EPID-based pre-treatment dose verification can be performed on a CP basis for VMAT plans. This model can measure pre-treatment doses for both flattened and unflattened beams in a time dependent manner which highlights deviations that are missed in integrated field verifications.Keywords
This publication has 38 references indexed in Scilit:
- A method for removing arm backscatter from EPID imagesMedical Physics, 2013
- Effect of statistical fluctuation in Monte Carlo based photon beam dose calculation on gamma index evaluationPhysics in Medicine & Biology, 2013
- Implementation of EPID transit dosimetry based on a through-air dosimetry algorithmMedical Physics, 2011
- Volumetric modulated arc therapy is superior to conventional intensity modulated radiotherapy - a comparison among prostate cancer patients treated in an Australian centreRadiation Oncology, 2011
- Using an EPID for patient‐specific VMAT quality assuranceMedical Physics, 2011
- Current status and future perspective of flattening filter free photon beamsMedical Physics, 2011
- Dosimetric verification of IMAT delivery with a conventional EPID system and a commercial portal dose image prediction toolMedical Physics, 2009
- IMRT commissioning: Multiple institution planning and dosimetry comparisons, a report from AAPM Task Group 119Medical Physics, 2009
- Epid cine acquisition mode for in vivo dosimetry in dynamic arc radiation therapyNuclear Instruments and Methods in Physics Research Section B: Beam Interactions with Materials and Atoms, 2008
- Correction of pixel sensitivity variation and off‐axis response for amorphous silicon EPID dosimetryMedical Physics, 2005