Non-Enzymatic Electrochemical Detection of Urine Creatinine Using Cobalt-Gold Bimetallic Nanoparticles
- 3 June 2024
- journal article
- research article
- Published by The Electrochemical Society in Journal of the Electrochemical Society
- Vol. 171 (6) , 067504
Abstract
This work presents the development of a non-enzymatic electrochemical sensor for creatinine in a neutral medium using a cobalt-gold bimetallic nanoparticles modified platinum electrode. The voltammetric detection of creatinine in a neutral phosphate buffer was based on the formation of a soluble cobalt-creatinine complex. The sensor exhibited good selectivity and a detection limit (S/N = 3) of 2.25 mM with two linear ranges from 6.4 to 83.2 mM. The sensitivity of the sensor was 0.621 and 1.135 μA mM−1 cm−2 at lower (6.4–51.2 mM) and higher (51.2–83.2 mM) detection ranges, respectively. The sensor performance was validated using urine samples and creatinine spiked urine samples.Keywords
Funding Information
- Department of Biotechnology (DBT), Government of India (102/IFD/SAN/1555/2018-2019)
This publication has 42 references indexed in Scilit:
- Silver-manganese nanocomposite modified screen-printed carbon electrode in the fabrication of an electrochemical, disposable biosensor strip for cystic fibrosisMicrochimica Acta, 2022
- Fe2O3/polyaniline supramolecular nanocomposite: A receptor free sensor platform for the quantitative determination of serum creatinineAnalytica Chimica Acta, 2020
- A copper oxide-ionic liquid/reduced graphene oxide composite sensor enabled by digital dispensing: Non-enzymatic paper-based microfluidic determination of creatinine in human blood serumAnalytica Chimica Acta, 2019
- Electrochemical creatinine sensor based on a glassy carbon electrode modified with a molecularly imprinted polymer and a Ni@polyaniline nanocompositeMicrochimica Acta, 2016
- A New Route for the Enzymeless Trace Level Detection of Creatinine Based on Reduced Graphene Oxide/Silver Nanocomposite BiosensorElectroanalysis, 2016
- Development of a Creatinine Sensor Based on a Molecularly Imprinted Polymer‐Modified Sol‐Gel Film on Graphite ElectrodeElectroanalysis, 2008
- Capacitive Creatinine Sensor Based on a Photografted Molecularly Imprinted PolymerElectroanalysis, 2002
- Microfabricated amperometric creatine and creatinine biosensorsAnalytica Chimica Acta, 1996
- Application of hydride generation to the determination of trace concentrations of arsenic by capacitively coupled microwave plasmaAnalytica Chimica Acta, 1993
- Cobalt(II) complexes of creatinineTransition Metal Chemistry, 1984