Comparison of PRIMO Monte Carlo code and Eclipse treatment planning system in calculation of dosimetric parameters in brain cancer radiotherapy
Authors:
Open Access
- 31 October 2022
- journal article
- research article
- Published by VM Media Group sp. z o.o in Reports of Practical Oncology and Radiotherapy
- Vol. 27 (5) , 863-874
Abstract
Background: It is important to evaluate the dose calculated by treatment planning systems (TPSs) and dose distribution in tumor and organs at risk (OARs). The aim of this study is to compare dose calculated by the PRIMO Monte Carlo code and Eclipse TPS in radiotherapy of brain cancer patients. Materials and methods: PRIMO simulation code was used to simulate a Varian Clinac 600C linac. The simulations were validated for the linac by comparison of the simulation and measured results. In the case of brain cancer patients, the dosimetric parameters obtained by the PRIMO code were compared with those calculated by Eclipse TPS. Gamma function analysis with 3%, 3 mm criteria was utilized to compare the dose distributions. The evaluations were based on the dosimetric parameters for the planning target volume (PTV) and OAR including Dmin, Dmean, and Dmax, homogeneity index (HI), and conformity index (CI). Results: The gamma function analysis showed a 98% agreement between the results obtained by the PRIMO code and measurement for the percent depth dose (PDD) and dose profiles. The corresponding value in comparing the dosimetric parameters from PRIMO code and Eclipse TPS for the brain patients was 94%, on average. The results of the PRIMO simulation were in good agreement with the measured data and Eclipse TPS calculations. Conclusions: Based on the results of this study, the PRIMO code can be utilized to simulate a medical linac with good accuracy and to evaluate the accuracy of treatment plans for patients with brain cancer. Background: It is important to evaluate the dose calculated by treatment planning systems (TPSs) and dose distribution in tumor and organs at risk (OARs). The aim of this study is to compare dose calculated by the PRIMO Monte Carlo code and Eclipse TPS in radiotherapy of brain cancer patients. Materials and methods: PRIMO simulation code was used to simulate a Varian Clinac 600C linac. The simulations were validated for the linac by comparison of the simulation and measured results. In the case of brain cancer patients, the dosimetric parameters obtained by the PRIMO code were compared with those calculated by Eclipse TPS. Gamma function analysis with 3%, 3 mm criteria was utilized to compare the dose distributions. The evaluations were based on the dosimetric parameters for the planning target volume (PTV) and OAR including Dmin, Dmean, and Dmax, homogeneity index (HI), and conformity index (CI). Results: The gamma function analysis showed a 98% agreement between the results obtained by the PRIMO code and measurement for the percent depth dose (PDD) and dose profiles. The corresponding value in comparing the dosimetric parameters from PRIMO code and Eclipse TPS for the brain patients was 94%, on average. The results of the PRIMO simulation were in good agreement with the measured data and Eclipse TPS calculations. Conclusions: Based on the results of this study, the PRIMO code can be utilized to simulate a medical linac with good accuracy and to evaluate the accuracy of treatment plans for patients with brain cancer.Keywords
This publication has 26 references indexed in Scilit:
- PRIMO: A graphical environment for the Monte Carlo simulation of Varian and Elekta linacsStrahlentherapie und Onkologie, 2013
- Dosimetric comparison of intensity modulated radiotherapy isocentric field plans and field in field (FIF) forward plans in the treatment of breast cancerJournal of Medical Physics, 2013
- A combined approach of variance-reduction techniques for the efficient Monte Carlo simulation of linacsPhysics in Medicine & Biology, 2012
- Homogeneity Index: An objective tool for assessment of conformal radiation treatmentsJournal of Medical Physics, 2012
- APENELOPE‐based system for the automated Monte Carlo simulation of clinacs and voxelized geometries—application to far‐from‐axis fieldsMedical Physics, 2011
- Design and implementation of a water phantom for IMRT, arc therapy, and tomotherapy dose distribution measurementsMedical Physics, 2007
- Evaluation of the gamma dose distribution comparison methodMedical Physics, 2003
- An investigation of a new amorphous silicon electronic portal imaging device for transit dosimetryMedical Physics, 2002
- A technique for the quantitative evaluation of dose distributionsMedical Physics, 1998
- Reporting and analyzing dose distributions: A concept of equivalent uniform doseMedical Physics, 1997