Development of a Rapid and Simple Digestion Method of Freshwater Sediments for As, Cd, Cr, Cu, Pb, Fe, and Zn Determination by Inductively Coupled Plasma-Optical Emission Spectroscopy (ICP-OES): An Evaluation of Dilute Nitric Acid
- 27 February 2019
- journal article
- research article
- Published by Taylor & Francis in Soil and Sediment Contamination: An International Journal
- Vol. 28 (3) , 323-333
Abstract
In this study, a simple and rapid methodology based on the hot-plate digestion method using dilute nitric acid solution was used to extract trace metals (such as As, Cd, Cr, Cu, Pb, Fe, and Zn) from freshwater sediments. The concentrations of the elements were determined using inductively coupled plasma-optical emission spectrometry (ICP-OES). The factors (temperature, nitric acid concentration, and volume) affecting the digestion method were optimized using one-factor-at-a-time (OFAT) or univariate methodology, and the optimization process was carried out using freshwater sediment certified reference material (CRM015). The optimal conditions for temperature, nitric acid concentration, and time in the method were 180°C, 10 mL of 5 mol L−1 HNO3, and 45 min, respectively. Under optimum conditions, the limit of detection (LOD) ranged between 0.02 and 0.08 µg L−1 and the limit of quantification (LOQ) ranged from 0.07 and 0.27 µg L−1. In addition, the method detection limits (MDLs) and method quantification limits (MQLs) were 0.10–0.17 and 0.30−0.57 µg g−1, respectively. The overall accuracy of the method determined by recovery of the trace elements in the CRMs ranged from 98 to 111% with the precision ranging from 1.4 to 5.8%. The method was successfully applied for the determination of target metals from real freshwater sediment samples.Keywords
Funding Information
- South African Agency for Science and Technology Advancement (99270)
This publication has 30 references indexed in Scilit:
- Mercury fractionation in dourada (Brachyplatystoma rousseauxii) of the Madeira River in Brazil using metalloproteomic strategiesTalanta, 2015
- Ligandless ultrasonic-assisted and ionic liquid-based dispersive liquid–liquid microextraction of copper, nickel and lead in different food samplesFood Chemistry, 2015
- Heavy Metal Polluted Soils: Effect on Plants and Bioremediation MethodsApplied and Environmental Soil Science, 2014
- Evaluation of a digestion procedure based on the use of diluted nitric acid solutions and H2O2for the multielement determination of whole milk powder and bovine liver by ICP-based techniquesJournal of Analytical Atomic Spectrometry, 2013
- Determination of 25 Trace Element Concentrations in Biological Reference Materials by ICP-MS following Different Microwave-Assisted Acid Digestion Methods Based on Scaling Masses of Digested SamplesInternational Scholarly Research Notices, 2013
- Understanding the process of microwave-assisted digestion combining diluted nitric acid and oxygen as auxiliary reagentMicrochemical Journal, 2011
- Remediation technologies for heavy metal contaminated groundwaterJournal of Environmental Management, 2011
- Heavy Metal Contamination in Soil, Water and Fodder and their Presence in Livestock and Products : A ReviewJournal of Environmental Science and Technology, 2011
- Determination of Cd, Cu, Pb and Zn in environmental samples: microwave-assisted total digestion versus aqua regia and nitric acid extractionAnalytica Chimica Acta, 2002
- Aufschlu biologischer Matrices f r die Bestimmung sehr niedriger Spurenelementgehalte bei begrenzter Einwaage mit Salpeters ure unter Druck in einem TeflongefAnalytical and Bioanalytical Chemistry, 1972