The effect of load on heat production, thermal effects and expenditure of time during implant site preparation – an experimental ex vivo comparison between piezosurgery and conventional drilling
- 27 November 2012
- journal article
- Published by Wiley in Clinical Oral Implants Research
- Vol. 25 (2) , e140-e148
Abstract
Piezoelectric surgery (PS) is meant to be a gentle osteotomy method. The aim of this study was to compare piezosurgical vs. conventional drilling methods for implant site preparation (ISP) - focusing on load-dependent thermal effect on hard tissue and the expenditure of ISP time. Three hundred and sixty ISP were performed on ex vivo pig heads using piezosurgery, spiral burs (SB) and trephine burs (TB). The load applied was increased from 0 to 1000 g in 100-g intervals. Temperature within the bone was measured with a thermocouple, and duration was recorded with a stop watch. Thermal effects were histomorphometrically analysed. Twelve ISPs per technique were performed at the lateral wall of the maxillary sinus. PS yields the highest mean temperatures (48.6 ± 3.4°C) and thermal effects (200.7 ± 44.4 μm), both at 900-1000 g. Duration is reduced with a plus of load and significantly longer in either case for PS (P < 0.05). There is a correlation of the applied load with all other examined factors for PS and TB. Temperature and histological effects decrease for SB beyond 500 g. PS yields significantly higher temperatures and thermal tissue alterations on load levels higher than 500 g and is significantly slower for ISP compared to SB and TB. For ISP with PS, a maximum load of 400 g should be maintained.Keywords
This publication has 36 references indexed in Scilit:
- An animal model for sinus floor elevation with great elevation heights. Macroscopic, microscopic, radiological and micro-CT analysis: ex vivoClinical Oral Implants Research, 2010
- Relevant Anatomic and Biomechanical Studies for Implant Possibilities on the Atrophic Maxilla: Critical Appraisal and Literature ReviewJournal of Prosthodontics, 2010
- Piezoosteotomy in orthognathic surgery versus conventional saw and chisel osteotomyOral and Maxillofacial Surgery, 2008
- Micromorphometrical analysis of conventional osteotomy techniques and ultrasonic osteotomy at the rabbit skullClinical Oral Implants Research, 2008
- Ultraschallbasiertes Knochenschneiden in der Oralchirurgie: eine Übersicht anhand von 60 PatientenfällenUltraschall in der Medizin - European Journal of Ultrasound, 2008
- Auflichtmikroskopische Untersuchungen an der Kaninchenkalotte nach ultraschallgestützter und konventioneller Osteotomie / Light microscopic examination of rabbit skulls following conventional and Piezosurgery®osteotomyBiomedical Engineering / Biomedizinische Technik, 2007
- Cytokines and Growth Factors Involved in the Osseointegration of Oral Titanium Implants Positioned Using Piezoelectric Bone Surgery Versus a Drill Technique: A Pilot Study in MinipigsJournal of Periodontology, 2007
- Increase of temperature during osteotomy.International Journal of Oral and Maxillofacial Surgery, 1991
- A Comparison of the Healing Rate of Bone after the Production of Defects by Various Rotary InstrumentsJournal of Dental Research, 1964
- Structural changes around nails and screws in human bonesThe Journal of Pathology and Bacteriology, 1953