Navigated femoral anteversion measurements: general precision and registration options
- 9 January 2009
- journal article
- research article
- Published by Springer Nature in Archives of orthopaedic and trauma surgery
- Vol. 129 (5) , 671-677
Abstract
Introduction Intraoperative reduction of femoral fractures can result in rotational malalignment. Navigation modules allow fluoroscopy-based femoral anteversion (AV) measurements; however, their clinical feasibility has not been fully evaluated. An important technical consideration when obtaining navigated femoral rotational alignment is the necessity for orthogonal navigated fluoroscopic images. Methods The current investigators hypothesized that there would be a critical imaging angle between the fluoroscopic images which optimized the acquisition of accurate rotational measurements. Following initial testing in plastic femora, 14 intact human cadaveric femora were subjected to CT and navigated fluoroscopic assessment of AV. The navigated registration process included the following test series: Standard perpendicular AP and lateral imaging at a 90° angle; planar rotation of the lateral fluoroscopy position perpendicular to the axis of femoral neck and angled to the AP position at 75°, 60°, and 50°; inlet and outlet imaging of the axis of the femoral neck at angles of 10°, 20°, and 30°. For all setups, the difference between the AV angle of the navigation system and the CT-measured angle was calculated. Results Results revealed no relevant differences between CT and navigated determination of plastic femora (1.5°). No significant deviations of cadaver femora with 90° angle imaging (mean 2.00°, range 0.00°–4.33°), 75° angle imaging (mean 1.95°, range 0.00°–3.33°) and 60° imaging (mean 2.00°, range 0.00°–3.33°) were found. However, significant deviations were found for the navigated 50° angled imaging technique (mean 5.02°, range 2.33°–7.67°). Conclusion These data demonstrated that the femoral AV angle can be calculated accurately by using two navigated fluoroscopic images when a minimum angle of 60° is utilized between the angled fluoroscopic images for the registration process. A difference of <60° does not allow for accurate navigated measurements.Keywords
This publication has 17 references indexed in Scilit:
- Computer Navigation Allows for Accurate Reduction of Femoral FracturesClinical Orthopaedics and Related Research, 2007
- Experimental validation of noninvasive referencing in navigated procedures on long bonesJournal of Orthopaedic Research, 2006
- Navigated intraoperative analysis of lower limb alignmentArchives of orthopaedic and trauma surgery, 2005
- Computerised navigation for closed reduction during femoral intramedullary nailingInjury, 2005
- Avoiding rotational malalignment after fractures of the femur by using the profile of the lesser trochanter: an in vitro studyArchives of orthopaedic and trauma surgery, 2005
- Computer aided long bone fracture treatmentInjury, 2004
- Intraoperative control of axes, rotation and length in femoral and tibial fractures technical noteInjury, 1998
- Femoral Torsion after Interlocked Nailing of Unstable Femoral FracturesThe Journal of Trauma: Injury, Infection, and Critical Care, 1995
- Die computertomographische AntetorsionswinkelbestimmungRöFo - Fortschritte auf dem Gebiet der Röntgenstrahlen und der bildgebenden Verfahren, 1986
- Anatomy of the Femoral Neck and Head, with Comparative Data From Caucasians and Hong Kong ChineseClinical Orthopaedics and Related Research, 1980