The Coefficient of Variation of Step Time Can Overestimate Gait Abnormality: Test-Retest Reliability of Gait-Related Parameters Obtained with a Tri-Axial Accelerometer in Healthy Subjects
Open Access
Abstract
The aim of this study was to investigate whether variation in gait-related parameters among healthy participants could help detect gait abnormalities. In total, 36 participants (21 men, 15 women; mean age, 35.7 ± 9.9 years) performed a 10-m walk six times while wearing a tri-axial accelerometer fixed at the L3 level. A second walk was performed ≥1 month after the first (mean interval, 49.6 ± 7.6 days). From each 10-m data set, the following nine gait-related parameters were automatically calculated: assessment time, number of steps, stride time, cadence, ground force reaction, step time, coefficient of variation (CV) of step time, velocity, and step length. Six repeated measurement values were averaged for each gait parameter. In addition, for each gait parameter, the difference between the first and second assessments was statistically examined, and the intraclass correlation coefficient (ICC) was calculated with the level of significance set at p < 0.05. Only the CV of step time showed a significant difference between the first and second assessments (p = 0.0188). The CV of step time also showed the lowest ICC, at <0.50 (0.425), among all parameters. Test–retest results of gait assessment using a tri-axial accelerometer showed sufficient reproducibility in terms of the clinical evaluation of all parameters except the CV of step time.Keywords
This publication has 32 references indexed in Scilit:
- Evaluation of gait symmetry using a tri-axial accelerometer in stroke patientsNeuroRehabilitation, 2018
- Intra-rater repeatability of gait parameters in healthy adults during self-paced treadmill-based virtual reality walkingComputer Methods in Biomechanics and Biomedical Engineering, 2017
- Accelerometer-based determination of gait variability in older adults with knee osteoarthritisGait & Posture, 2016
- Test-Retest Reliability of an Automated Infrared-Assisted Trunk Accelerometer-Based Gait Analysis SystemSensors, 2016
- Validity and reliability of a portable gait analysis system for measuring spatiotemporal gait characteristics: comparison to an instrumented treadmillJournal of NeuroEngineering and Rehabilitation, 2016
- Mobility and cognitive correlates of dual task cost of walking in persons with multiple sclerosisDisability and Rehabilitation, 2013
- Objective assessment of motor fatigue in multiple sclerosis using kinematic gait analysis: a pilot studyJournal of NeuroEngineering and Rehabilitation, 2011
- Assessment of gait in subcortical vascular encephalopathy by computerized analysis: a cross-sectional and longitudinal studyJournal of Neurology, 2000
- Foot-force measuring device for clinical assessment of pathological gaitMedical & Biological Engineering & Computing, 1978
- The assessment of gait.1969