Impact of x-ray contamination in the dose monitor on beam abortion during uninterrupted continuous delivery in proton therapy
- 5 June 2026
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 71 (13)
Abstract
Objective: Real-time Gated Proton Therapy (RGPT) enables precise irradiation of moving targets and is an important motion-management technique in proton therapy. Uninterrupted continuous delivery (UCD), in which fluoroscopic X-rays remain on during proton irradiation, may improve beam stability and treatment efficiency, particularly in synchrotronbased systems using continuous beams. However, scattered fluoroscopic X-rays may be detected by the dose monitor (DM) and mistakenly counted as proton monitor units, potentially causing beam aborts. This study aimed to quantify scattered fluoroscopic X-rays detected by the DM during RGPT and evaluate their impact on beam delivery during simultaneous fluoroscopy and proton irradiation.Methods: Scattered fluoroscopic X-ray signals detected by the DM were measured using an RGPT system. Phantom experiments evaluated scattered X-ray dose rates while varying phantom thickness, fluoroscopic parameters, and range-shifter conditions. Beam-abort simulations were conducted under UCD conditions for both phantom and clinical liver cases.Abort probability was assessed as a function of proton MU per spot, scattered X-ray MU per frame, and fluoroscopy frame rate (FPS).Results: DM signals from scattered fluoroscopic X-rays increased with tube voltage, tube current, field size, and phantom thickness, and were higher without a range shifter. The maximum dose rate was 0.0211 MU/frame at 125 kV, 80 mAs, 200 × 200 mm², and PW-30 cm without a range shifter, representing an 11.7-fold increase compared with in-air conditions. Beam-abort simulations showed abort proportions below 2% for monoenergetic fields and below 3% for clinical liver plans across all tested conditions. Significance: With the DM positioned closer to the X-ray tube, scattered fluoroscopic X-ray contamination increased with phantom thickness, differing from prior reports. Optimization of proton dose rate, imaging parameters, and system geometry can minimize beam interruptions, supporting the clinical feasibility of UCD.Keywords
Funding Information
- Japan Society for the Promotion of Science (22K07695)
- Japan Society for the Promotion of Science (25K11008)
- Duke-NUS Oncology Academic Program Goh Foundation Proton Research Programme (08/FY2025/EX(SL)/ 000364)
This publication has 23 references indexed in Scilit:
- Dosimetric impact and safety of uninterrupted fluoroscopic‐gated proton therapyMedical Physics, 2025
- Characterization of secondary‐radiation background in X‐ray flat‐panel detectors during scanning proton beam irradiationMedical Physics, 2025
- Optimizing dose rate to minimize delivery time in proton pencil beam dose‐driven continuous scanningMedical Physics, 2025
- Scanned particle‐beam tracking with beam correction based on predictive volumetric imaging: A simulation studyMedical Physics, 2025
- Quality assurance of the inter‐spot beam delivery of a dose driven continuous scanning proton therapy systemMedical Physics, 2025
- Dosimetric study of synchrotron rapid beam off control and skip spot function for high beam intensity proton therapyMedical Physics, 2024
- Real-time gated proton therapy with a reduced source to imager distance: Commissioning and quality assurancePhysica Medica, 2024
- Beam delivery characteristics of the Hitachi carbon ion scanning system at Osaka Heavy Ion Medical Accelerator in Kansai (HIMAK)Medical Physics, 2023
- The effect of spill change on reliable absolute dosimetry in a synchrotron proton spot scanning systemMedical Physics, 2023
- AAPM Task Group Report 290: Respiratory motion management for particle therapyMedical Physics, 2022