Functional forms for photon spectra of clinical linacs
- 29 November 2011
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 57 (1) , 31-50
Abstract
Specifying photon spectra of clinical linacs using a functional form is useful for many applications, including virtual source modelling and spectral unfolding from dosimetric measurements such as transmission data or depth-dose curves. In this study, 11 functional forms from the literature are compiled and quantitatively compared. A new function is proposed which offers improvements over existing ones. The proposed function is simple, physics-based and has four free parameters, one of which is the mean incident electron kinetic energy. A comprehensive benchmark set of validated, high-precision Monte Carlo spectra is generated and used to evaluate the strengths and limitations of different functions. The benchmark set has 65 spectra (3.5-30 MV) from Varian, Elekta, Siemens, Tomotherapy, Cyberknife and research linacs. The set includes spectra on- and off-axis from linacs with and without a flattening filter, and in treatment and imaging modes. The proposed function gives the lowest spectral deviations among all functions. It reproduces the energy fluence values in each bin for the benchmark set with a normalized root-mean-square deviation of 1.7%. The mean incident electron kinetic energy, maximum photon energy, most-probable energy and average energy are reproduced, on average, within 1.4%, 4.3%, 3.9% and 0.6% of their true values, respectively. The proposed function is well behaved when used for spectral unfolding from dosimetric data. The contribution of the 511 keV annihilation peak and the energy spread of the incident electron beam can be added as additional free parameters.Keywords
This publication has 38 references indexed in Scilit:
- TH-E-BRC-09: Beyond Self-Consistency in Beam Commissioning: Determination of True Linac SpectraMedical Physics, 2011
- Benchmarking of Monte Carlo simulation of bremsstrahlung from thick targets at radiotherapy energiesMedical Physics, 2008
- A flexible Monte Carlo tool for patient or phantom specific calculations: comparison with preliminary validation measurementsJournal of Physics: Conference Series, 2008
- Monte Carlo study of a Cyberknife stereotactic radiosurgery systemMedical Physics, 2006
- Reference photon dosimetry data and reference phase space data for the photon beam from Varian Clinac 2100 series linear acceleratorsMedical Physics, 2004
- Enhanced spectral discrimination through the exploitation of interface effects in photon dose dataMedical Physics, 2004
- Extraction of the photon spectra from measured beam parametersMedical Physics, 1998
- Reconstruction of 6 MV photon spectra from measured transmission including maximum energy estimationPhysics in Medicine & Biology, 1997
- Bremsstrahlung review: An analysis of the Schiff spectrumMedical Physics, 1991
- Forward‐directed bremsstrahlung of 10‐ to 30‐MeV electrons incident on thick targets of Al and PbMedical Physics, 1990