Evaluating agreement between separate capillary sampling sites and venous concentrations of β-hydroxybutyrate measured by a point-of-care device and liquid chromatography tandem mass spectrometry
- 2 January 2025
- journal article
- Published by Taylor & Francis in Scandinavian Journal of Clinical and Laboratory Investigation
- Vol. 85 (1) , 58-63
Abstract
β-hydroxybutyrate (BHB) is recommended as a measure of ketosis and is often assessed by capillary samples on point-of-care (POC) meters. However, liquid chromatography tandem mass spectrometry (LC-MS/MS) is considered the gold standard for assessing venous samples. The POC device KetoSureTM is recommended only for capillary samples from finger pricks. So far, KetoSureTM has not been compared to LC-MS/MS just as it has not been evaluated if the sampling site influences the BHB concentration. Blood samples were collected from 16 healthy, fasting individuals before and after ingestion of ketone monoester. BHB concentrations were measured in capillary samples from the earlobe and fingertip, and in venous blood using KetoSureTM. Venous plasma samples were collected for BHB quantification using LC-MS/MS. No sign of significant difference between values of BHB measured from the two capillary sampling sites were found. Interestingly, significantly higher values of BHB were measured in capillary samples compared to venous samples reflecting a systematic proportional relationship. No systematic difference was observed in the measured BHB concentrations when comparing KetoSureTM and LC-MS/MS results: However, a significant mean bias of 32% reflects a skewness at very low BHB concentrations. In conclusion, capillary BHB concentration did not exhibit variation between the earlobe and fingertip. Conversely, a significant bias was observed between venous and capillary blood and between the POC and LC-MS/MS methods. It is recommended that caution be exercised if individual monitoring of BHB changes encompasses both capillary and venous sampling.Keywords
This publication has 30 references indexed in Scilit:
- Nutrient sensing: LEAP2 concentration in response to fasting, glucose, lactate, and β-hydroxybutyrate in healthy young malesThe American Journal of Clinical Nutrition, 2023
- Ketone monoester increases circulating levels of LEAP2 and decreases appetite in healthy menDiabetes, Obesity and Metabolism, 2023
- Evaluation of the point-of-care devices KetoSureTM and StatStrip Express® blood ketone tests using β-hydroxybutyrate spiked samplesScandinavian Journal of Clinical and Laboratory Investigation, 2022
- Evaluation of Nova StatStrip and FreeStyle Precision Pro blood ketone tests using 3‐hydroxybutyrate doped samplesJournal of Clinical Laboratory Analysis, 2019
- Multicentric evaluation of eight glucose and four ketone blood metersClinical Biochemistry, 2015
- Simultaneous determination of β-hydroxybutyrate and β-hydroxy-β-methylbutyrate in human whole blood using hydrophilic interaction liquid chromatography electrospray tandem mass spectrometryClinical Biochemistry, 2013
- Limitations to using point of care blood ketone testing to monitor DKA treatmentPractical Diabetes, 2013
- Blood β‐hydroxybutyrate vs. urine acetoacetate testing for the prevention and management of ketoacidosis in Type 1 diabetes: a systematic reviewDiabetic Medicine, 2013
- Urinary acetoacetate or capillary ??-hydroxybutyrate for the diagnosis of ketoacidosis in the Emergency Department settingEuropean Journal of Emergency Medicine, 2004
- Screening for Ketonemia in Patients With DiabetesAnnals of Emergency Medicine, 1999