Comparison of two catheters measuring local impedance: local impedance variation vs lesion characteristics and steam pops

Authors:

Abstract
Purpose The size of the distal electrode and the method of measuring local impedance (LI) are different between the IntellaNav MiFi-OI™ (MiFi-OI) and IntellaNav STABLE POINT™ (SP) catheters. We investigated the impact of these differences on LI, efficacy, and safety of radiofrequency (RF) applications. Methods RF applications at a range of powers (30 W, 40 W, and 50 W), contact forces (10 g and 20 g), and durations (10–120 s) were performed in excised porcine hearts (N = 48). LI variation was defined by δLI-drop (= initial LI − post-RF LI) and %LI-drop (= δLI-drop/initial LI) × 100, and the relationship between lesion characteristics and LI variation was compared. Results A total of 576 lesions were examined. Although absolute δLI-drop during RF applications was significantly larger for the SP than MiFi-OI catheter (47[31–65]ohm for SP vs 37[24–51]ohm for MiFi-OI, p < 0.0001), %LI-drop was similar (23.3 [15.5–30.6]% in SP vs 24.9[17.3–32.5]% in MiFi-OI, p = 0.10). Although lesions produced by both catheters were similarly correlated with LI variation, the SP catheter produced generally larger lesions (depth; 5.0 [3.7–6.1]mm vs 4.7 [3.3–6.0]mm, p = 0.06; surface areas, 46.9 [36.8–58.8]mm2 vs 44.7 [34.3–55.5]mm2, p = 0.02; volume, 321 [165–533]mm3 vs 265[141–471]mm3, p = 0.02). Steam pops were similarly observed with both catheters. In both catheters, %LI-drop was superior to δLI-drop in correlation to lesion size (p < 0.0001) and in predicting steam pops (p < 0.01). Conclusions Although no difference in safety profile is observed between MiFi-OI and SP catheters, the SP catheter produces larger lesions. %LI-drop is superior to δLI-drop in correlation to lesion size and in predicting steam pops as well as in normalizing the difference between catheters.

This publication has 24 references indexed in Scilit: