Comparison of two catheters measuring local impedance: local impedance variation vs lesion characteristics and steam pops
- 19 April 2022
- journal article
- research article
- Published by Springer Nature in Journal of Interventional Cardiac Electrophysiology
- Vol. 65 (2) , 419-428
Abstract
Purpose The size of the distal electrode and the method of measuring local impedance (LI) are different between the IntellaNav MiFi-OI™ (MiFi-OI) and IntellaNav STABLE POINT™ (SP) catheters. We investigated the impact of these differences on LI, efficacy, and safety of radiofrequency (RF) applications. Methods RF applications at a range of powers (30 W, 40 W, and 50 W), contact forces (10 g and 20 g), and durations (10–120 s) were performed in excised porcine hearts (N = 48). LI variation was defined by δLI-drop (= initial LI − post-RF LI) and %LI-drop (= δLI-drop/initial LI) × 100, and the relationship between lesion characteristics and LI variation was compared. Results A total of 576 lesions were examined. Although absolute δLI-drop during RF applications was significantly larger for the SP than MiFi-OI catheter (47[31–65]ohm for SP vs 37[24–51]ohm for MiFi-OI, p < 0.0001), %LI-drop was similar (23.3 [15.5–30.6]% in SP vs 24.9[17.3–32.5]% in MiFi-OI, p = 0.10). Although lesions produced by both catheters were similarly correlated with LI variation, the SP catheter produced generally larger lesions (depth; 5.0 [3.7–6.1]mm vs 4.7 [3.3–6.0]mm, p = 0.06; surface areas, 46.9 [36.8–58.8]mm2 vs 44.7 [34.3–55.5]mm2, p = 0.02; volume, 321 [165–533]mm3 vs 265[141–471]mm3, p = 0.02). Steam pops were similarly observed with both catheters. In both catheters, %LI-drop was superior to δLI-drop in correlation to lesion size (p < 0.0001) and in predicting steam pops (p < 0.01). Conclusions Although no difference in safety profile is observed between MiFi-OI and SP catheters, the SP catheter produces larger lesions. %LI-drop is superior to δLI-drop in correlation to lesion size and in predicting steam pops as well as in normalizing the difference between catheters.Keywords
This publication has 24 references indexed in Scilit:
- Lesion Size and Safety Comparison Between the Novel Flex Tip on the FlexAbility Ablation Catheter and the Solid Tips on the ThermoCool and ThermoCool SF Ablation CathetersJournal of Cardiovascular Electrophysiology, 2015
- Relationship Between Catheter Contact Force and Radiofrequency Lesion Size and Incidence of Steam Pop in the Beating Canine HeartCirculation: Arrhythmia and Electrophysiology, 2014
- Clinical experience with a novel, irrigated, flexible tip ablation catheter in atrial fibrillation ablation.2014
- Effects of Open‐Irrigated Radiofrequency Ablation Catheter Design on Lesion Formation and Complications: In Vitro Comparison of 6 Different DevicesJournal of Cardiovascular Electrophysiology, 2013
- Importance of Catheter Contact Force During Irrigated Radiofrequency Ablation: Evaluation in a Porcine Ex Vivo Model Using a Force‐Sensing CatheterJournal of Cardiovascular Electrophysiology, 2010
- Novel Contact Force Sensor Incorporated in Irrigated Radiofrequency Ablation Catheter Predicts Lesion Size and Incidence of Steam Pop and ThrombusCirculation: Arrhythmia and Electrophysiology, 2008
- Steam pops during irrigated radiofrequency ablation: Feasibility of impedance monitoring for preventionHeart Rhythm, 2008
- Tissue Temperatures and Lesion Size During Irrigated Tip Catheter Radiofrequency Ablation: An In Vitro Comparison of Temperature‐Controlled Irrigated Tip Ablation, Power‐Controlled Irrigated Tip Ablation, and Standard Temperature‐Controlled AblationPacing and Clinical Electrophysiology, 2000
- The Effects of Electrode‐Tissue Contact on Radiofrequency Lesion GenerationPacing and Clinical Electrophysiology, 1997
- Impedance Monitoring During Radiofrequency Catheter Ablation in HumansPacing and Clinical Electrophysiology, 1992