Dosimetric characteristics of electron beams produced by a mobile accelerator for IORT
Authors:
- M PimpinellaMPM Pimpinella
- D MihailescuDMD Mihailescu
- A S GuerraAGA S Guerra
- R F LaitanoRLR F Laitano
- 1 October 2007
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 52 (20) , 6197-6214
Abstract
Energy and angular distributions of electron beams with different energies were simulated by Monte Carlo calculations. These beams were generated by the NOVAC7 system (Hitesys, Italy), a mobile electron accelerator specifically dedicated to intra-operative radiation therapy (IORT). The electron beam simulations were verified by comparing the measured dose distributions with the corresponding calculated distributions. As expected, a considerable difference was observed in the energy and angular distributions between the IORT beams studied in the present work and the electron beams produced by conventional accelerators for non-IORT applications. It was also found that significant differences exist between the IORT beams used in this work and other IORT beams with different collimation systems. For example, the contribution from the scattered electrons to the total dose was found to be up to 15% higher in the NOVAC7 beams. The water-to-air stopping power ratios of the IORT beams used in this work were calculated on the basis of the beam energy distributions obtained by the Monte Carlo simulations. These calculated stopping power ratios, s(w,air), were compared with the corresponding s(w,air) values recommended by the TRS-381 and TRS-398 IAEA dosimetry protocols in order to estimate the deviations between a dosimetry based on generic parameters and a dosimetry based on parameters specifically obtained for the actual IORT beams. The deviations in the s(w,air) values were found to be as large as up to about 1%. Therefore, we recommend that a preliminary analysis should always be made when dealing with IORT beams in order to assess to what extent the possible differences in the s(w,air) values have to be accounted for or may be neglected on the basis of the specific accuracy needed in clinical dosimetry.Keywords
This publication has 9 references indexed in Scilit:
- Measurements of output factors with different detector types and Monte Carlo calculations of stopping-power ratios for degraded electron beamsPhysics in Medicine & Biology, 2004
- Dosimetry characteristics of degraded electron beams investigated by Monte Carlo calculations in a setup for intraoperative radiation therapyPhysics in Medicine & Biology, 2002
- Design and dosimetry characteristics of a soft-docking system for intraoperative radiation therapyInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2000
- Accurate condensed history Monte Carlo simulation of electron transport. I. EGSnrc, the new EGS4 versionMedical Physics, 2000
- AAPM's TG‐51 protocol for clinical reference dosimetry of high‐energy photon and electron beamsMedical Physics, 1999
- R50 as a beam quality specifier for selecting stopping‐power ratios and reference depths for electron dosimetryMedical Physics, 1996
- Calculation of stopping‐power ratios using realistic clinical electron beamsMedical Physics, 1995
- BEAM: A Monte Carlo code to simulate radiotherapy treatment unitsMedical Physics, 1995
- Influence of energy and angular spread on stopping-power ratios for electron beamsPhysics in Medicine & Biology, 1989