Artefact and ablation performance of an MR-conditional high-power microwave system in bovine livers: an ex vivo study
Open Access
- 23 September 2019
- journal article
- research article
- Published by Springer Nature in European Radiology Experimental
- Vol. 3 (1) , 1-9
Abstract
Background: We evaluated a magnetic resonance (MR)-conditional high-power microwave ablation system. Methods: An exvivo 1.5-T evaluation was conducted by varying the sequence (T1-weighted volume interpolated breath-hold examination, T1w-VIBE; T1-weighted fast low-angle shot, T1w-FLASH; T2-weighted turbo spin-echo, T2w-TSE), applicator angulation to B0 (A-to-B0), slice orientation, and encoding direction. Tip location error (TLE) and artefact diameters were measured, and influence of imaging parameters was assessed with analysis of variance and post hoc testing. Twenty-four exvivo ablations were conducted in three bovine livers at 80 W and 120 W. Ablation durations were 5, 10, and 15 min. Ablation zones were compared for short-axis diameter (SAD), volume, and sphericity index (SI) with unpaired t test. Results: The artefact pattern was similar for all sequences. The shaft artefact (4.4 ± 2.9 mm, mean ± standard deviation) was dependent on the sequence (p = 0.012) and the A-to-B0 (p < 0.001); the largest shaft diameter was measured with T1w-FLASH (6.3 ± 3.4 mm) and with perpendicular A-to-B0 (6.7 ± 2.4 mm). The tip artefact (1.6 ± 0.7 mm) was dependent on A-to-B0 (p = 0.001); TLE was -2.6 ± 1.0 mm. Ablation results at the maximum setting (15 min, 120 W) were SAD = 42.0 ± 1.41 mm; volume = 56.78 ± 3.08 cm3, SI = 0.68 ± 0.05. In all ablations, SI ranged 0.68–0.75 with the smallest SI at 15 min and 120 W (p = 0.048). Conclusion: The system produced sufficiently large ablation zones and the artefact was appropriate for MR-guided interventions.Keywords
This publication has 35 references indexed in Scilit:
- MR-guided radiofrequency ablation using a wide-bore 1.5-T MR system: clinical results of 213 treated liver lesionsEuropean Radiology, 2012
- Comparison of Microwave Ablation and Multipolar Radiofrequency Ablation In Vivo Using Two Internally Cooled ProbesAmerican Journal of Roentgenology, 2012
- Performing MR-guided biopsies in clinical routine: factors that influence accuracy and procedure timeEuropean Radiology, 2011
- Image-Based Monitoring of Magnetic Resonance-Guided Thermoablative Therapies for Liver TumorsCardioVascular and Interventional Radiology, 2011
- In Vitro Assessment of Needle Artifacts with an Interactive Three-dimensional MR Fluoroscopy SystemJournal of Vascular and Interventional Radiology, 2010
- Radiofrequency ablation of small liver malignancies under magnetic resonance guidance: progress in targeting and preliminary observations with temperature monitoringEuropean Radiology, 2009
- Evaluation of Magnetic Resonance Imaging-Compatible Needles and Interactive Sequences for Musculoskeletal Interventions Using an Open High-Field Magnetic Resonance Imaging ScannerCardioVascular and Interventional Radiology, 2009
- Radiofrequency Ablation: Simultaneous Application of Multiple Electrodes via Switching Creates Larger, More Confluent Ablations than Sequential Application in a Large Animal ModelJournal of Vascular and Interventional Radiology, 2009
- Adjusting for multiple testing—when and how?Journal of Clinical Epidemiology, 2001
- Multiple testing in clinical trialsStatistics in Medicine, 1991