The delivered dose assessment in pancreas SBRT with the target position determined using an in-house position monitoring system
Authors:
Open Access
- 28 November 2022
- journal article
- research article
- Published by Frontiers Media SA in Frontiers in Oncology
- Vol. 12, 1009916
Abstract
Purpose: This study assessed the delivered dose accuracy in pancreas SBRT by incorporating the real-time target position determined using an in-house position monitoring system.Methods and materials: An online image-based position monitoring system, SeedTracker, was developed to monitor radiopaque marker positions using monoscopic x-ray images, available from the Elekta XVI imaging system. This system was applied to patients receiving SBRT for pancreatic cancer on the MASTERPLAN Pilot trial (ACTRN 12617001642370). All patients were implanted pre-treatment with at least three peri-tumoral radiopaque markers for target localisation. During treatment delivery, marker positions were compared to expected positions delineated from the planning CT. The position tolerance of ±3mm from the expected position of the markers was set to trigger a gating event (GE) during treatment. The dosimetric impact of position deviations and actual dose delivered with position corrections was assessed by convolving the plan control point dose matrices with temporal target positions determined during treatment.Results: Eight patients were treated within this study. At least one GE was observed in 38% of the treatment fractions and more than one GE was observed in 10% of the fractions. The position deviations resulted in the mean(range) difference of -0.1(-1.1 - 0.4)Gy in minimum dose to tumour and 1.9(-0.1- 4.6)Gy increase to Dmax to duodenum compared to planned dose. In actual treatment delivery with the patient realignment, the mean difference of tumour min dose and duodenal Dmax was reduced to 0.1(-1.0 – 1.1)Gy and 1.1 (-0.7 - 3.3)Gy respectively compared to the planned dose.Conclusions: The in-house real-time position monitoring system improved the treatment accuracy of pancreatic SBRT in a general-purpose linac and enabled assessment of delivered dose by incorporating the temporal target position during delivery. The intrafraction motion impacts the dose to tumour even if target position is maintained within a 3mm position tolerance.Keywords
This publication has 30 references indexed in Scilit:
- Stereotactic Body Radiation Therapy for Locally Advanced Pancreatic Cancer: A Systematic Review and Pooled Analysis of 19 TrialsInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2017
- Real time tracking in liver SBRT: comparison of CyberKnife and Vero by planning structure-basedγ-evaluation and dose-area-histogramsPhysics in Medicine & Biology, 2016
- An online x‐ray based position validation system for prostate hypofractionated radiotherapyMedical Physics, 2016
- Determination of Acquisition Frequency for Intrafractional Motion of Pancreas in CyberKnife RadiotherapyThe Scientific World Journal, 2014
- Outcomes with FOLFIRINOX for borderline resectable and locally unresectable pancreatic cancerJournal of Surgical Oncology, 2013
- Planning 4-Dimensional Computed Tomography (4DCT) Cannot Adequately Represent Daily Intrafractional Motion of Abdominal TumorsInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2013
- Predictor of Severe Gastroduodenal Toxicity After Stereotactic Body Radiotherapy for Abdominopelvic MalignanciesInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2012
- Characterization and Management of Interfractional Anatomic Changes for Pancreatic Cancer RadiotherapyInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2012
- Pancreatic Tumor Motion on a Single Planning 4D-CT Does Not Correlate With Intrafraction Tumor Motion During TreatmentAmerican Journal of Clinical Oncology, 2009
- Characterization of Pancreatic Tumor Motion Using Cine MRI: Surrogates for Tumor Position Should Be Used With CautionInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2009