Quantitative and qualitative evaluation of hybrid iterative reconstruction, with and without noise power spectrum models: A phantom study
Open Access
- 28 February 2018
- journal article
- research article
- Published by Wiley in Journal of Applied Clinical Medical Physics
- Vol. 19 (3) , 318-325
Abstract
The purpose of this phantom study was to investigate the feasibility of dose reduction with hybrid iterative reconstruction, with and without a noise power spectrum (NPS) model, using both quantitative and qualitative evaluations. Standard dose (SD), three‐quarter dose (TQD), and half‐dose (HD) of radiation were used. Images were reconstructed with filtered back projection (FBP), adaptive iterative dose reduction 3D (AIDR 3D) (MILD, STR), and AIDR 3D enhanced (eAIDR 3D) (eMILD, eSTR). An NPS analysis, task‐based modulation transfer function (MTFtask) analysis, and comparisons of low‐contrast detectability and image texture were performed. Although the eAIDR 3D had a higher NPS value in the high‐frequency range and improved image texture and resolution as compared with AIDR 3D at the same radiation dose and iteration levels, it yielded higher noise than AIDR 3D. Additionally, although there was no statistically significant difference between SD‐FBP and the TQD series in the comparison of the mean area under the curve (AUC), the mean AUC was statistically significantly different between SD‐FBP and the HD series. NPS values in the high‐frequency range, 10% MTFtask values, low‐contrast detectability, and image textures of TQD‐eMILD were comparable to those of SD‐FBP. Our findings suggested that using eMILD can reduce the radiation dose by 25%, while potentially maintaining diagnostic performance, spatial resolution, and image texture; this could support selecting the appropriate protocol in a clinical setting.Keywords
This publication has 40 references indexed in Scilit:
- Influence of Sinogram Affirmed Iterative Reconstruction of CT Data on Image Noise Characteristics and Low-Contrast Detectability: An Objective ApproachPLOS ONE, 2013
- Flexible Image Evaluation: iPad versus Secondary-class Monitors for Review of MR Spinal Emergency Cases, A Comparative StudyAcademic Radiology, 2012
- Image quality of Adaptive Iterative Dose Reduction 3D of coronary CT angiography of 640-slice CT: comparison with filtered back-projectionThe International Journal of Cardiovascular Imaging, 2012
- Model-Based Iterative Reconstruction Technique for Ultralow-Dose Computed Tomography of the LungInvestigative Radiology, 2012
- Spectral-Domain Optical Coherence Tomography with Multiple B-Scan Averaging for Enhanced Imaging of Retinal DiseasesOphthalmology, 2008
- Effect of sample size for normal database on diagnostic performance of brain FDG PET for the detection of Alzheimer's disease using automated image analysisNuclear Medicine Communications, 2008
- Application of the noise power spectrum in modern diagnostic MDCT: part I. Measurement of noise power spectra and noise equivalent quantaPhysics in Medicine & Biology, 2007
- MRI angiography is superior to helical CT for detection of HCC prior to liver transplantation: An explant correlationHepatology, 2003
- Mri Angiography Is Superior to Helical Ct for Detection of Hcc Prior to Liver Transplantation: An Explant CorrelationHepatology, 2003
- Use of CT Hounsfield unit density to identify ablated tumor after laparoscopic radiofrequency ablation of hepatic tumorsrid=""id=""Presented at the annual meeting of the Society of American Gastrointestinal Endoscopic Surgeons (SAGES), Atlanta, Georgia, USA, 29 March–1 April 2000Surgical Endoscopy, 2000