Bias and precision of SPECT-based 177Lu activity-concentration estimation using a ring-configured solid-state versus a dual-headed anger system
Open Access
- 4 November 2024
- journal article
- research article
- Published by Springer Nature in EJNMMI Physics
- Vol. 11 (1) , 1-21
Abstract
Background: The aim was to compare bias and precision for 177Lu-SPECT activity-concentration estimation using a dual-headed Anger SPECT system and a ring-configured CZT SPECT system. This was investigated for imaging at 208 keV and 113 keV, respectively. Methods: Phantom experiments were performed on a GE Discovery 670 system with 5/8’’ NaI(Tl) crystal (dual-headed Anger system) and a GE StarGuide (ring-configured CZT system). Six spheres (1.2 mL to 113 mL) in a NEMA PET body phantom were filled with 99mTc and 177Lu, separately. Mean relative errors and coefficients of variation (CV) in estimated sphere activity concentration were studied over six timeframes of 10 min each for the two systems. For 177Lu, similar acquisitions were also performed for an anthropomorphic phantom with two spheres (10 mL and 25 mL) in a liver with non-radioactive background and a sphere-to-background ratio of 15:1. Tomographic reconstruction was performed using OS-EM with 10 subsets with compensation for attenuation, scatter, and distance-dependent spatial resolution. For the Anger system, up to 40 iterations were used and for the ring-configured CZT system up to 30 iterations were used. Results: The two systems showed similar mean relative errors and CVs for 177Lu when using an energy window around 208 keV, while the ring-configured system demonstrated a lower bias for a similar CV compared to the Anger system for 99mTc and for 177Lu when using an energy window around 113 keV. However, total activity in the phantom tended to be overestimated in both systems for these cases. Conclusions: The ring-configured CZT system is a viable alternative to the dual-headed Anger system equipped with medium-energy collimators for 177Lu-SPECT and shows a potential advantage for activity-concentration estimation when operated at 113 keV. However, further consideration of the preservation of total activity is warranted.Keywords
Funding Information
- Fru Berta Kamprads Stiftelse (FBKS-2024-7-583)
- Cancerfonden (21 1754 Pj 01H)
This publication has 40 references indexed in Scilit:
- SPECT at the speed of PET: a feasibility study of CZT-based whole-body SPECT/CT in the post 177Lu-DOTATATE and 177Lu-PSMA617 settingEuropean Journal of Nuclear Medicine and Molecular Imaging, 2023
- Ultra-fast high resolution whole-body SPECT after treatment with 153Sm-EDTMP using 3D-ring CZT: applying new technology to an old tracerEuropean Journal of Nuclear Medicine and Molecular Imaging, 2022
- EANM dosimetry committee recommendations for dosimetry of 177Lu-labelled somatostatin-receptor- and PSMA-targeting ligandsEuropean Journal of Nuclear Medicine and Molecular Imaging, 2022
- Recent Developments in SPECT/CTSeminars in Nuclear Medicine, 2022
- Can Q.Clear reconstruction be used to improve [68 Ga]Ga-DOTANOC PET/CT image quality in overweight NEN patients?European Journal of Nuclear Medicine and Molecular Imaging, 2021
- High quality attenuation-corrected myocardial perfusion images may be achieved in patients with morbid obesity using a whole-body CZT camera and a low-dose injection protocolJournal of Nuclear Cardiology, 2020
- An energy-optimized collimator design for a CZT-based SPECT cameraNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2016
- Pharmacokinetic digital phantoms for accuracy assessment of image-based dosimetry in177Lu-DOTATATE peptide receptor radionuclide therapyPhysics in Medicine & Biology, 2015
- Scatter correction improves concordance in SPECT MPI with a dedicated cardiac SPECT solid-state cameraJournal of Nuclear Cardiology, 2015
- Scintillation CameraReview of Scientific Instruments, 1958