Investigation of the Power Consumption of the PETsys TOFPET2 ASIC
Open Access
- 21 November 2019
- journal article
- research article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Radiation and Plasma Medical Sciences
- Vol. 4 (3) , 378-388
Abstract
In state-of-the-art positron emission tomography (PET) systems, application-specific integrated circuits (ASICs) are commonly used to precisely digitize the signals of analog silicon photomultipliers (SiPMs). However, when operating PET electronics in a magnetic resonance (MR) system, one faces the challenge of mutual interference of these imaging techniques. To prevent signal deterioration along long analog signal lines, PET electronics with a low-power consumption digitizing the signals close to the SiPMs are preferred. In this article, we evaluate the power consumption of the TOFPET2 ASIC. Its power consumption ranges from 3.6 mW/channel to 7.2 mW/channel as a function of the input stage impedance and discriminator noise settings. We present an analytical model allowing to compute the power consumption of a given ASIC configuration. The input stage impedance and discriminator noise have an impact on the coincidence resolution time (CRT), energy resolution, and photon trigger level. Since the TOFPET2 ASIC delivers state-of-the-art performance with a power consumption similar or even lower than other ASICs typically used for PET applications, it is a favorable candidate to digitize the signals of SiPMs in future simultaneous PET/MR systems.Keywords
All Related Versions
Funding Information
- European Unions Horizon 2020 Research and Innovation Programme (667211)
This publication has 41 references indexed in Scilit:
- STIC3 – Silicon Photomultiplier Timing Chip with picosecond resolutionNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2015
- PET-MRI: a review of challenges and solutions in the development of integrated multimodality imagingPhysics in Medicine & Biology, 2015
- Update on Time-of-Flight PET ImagingJournal of Nuclear Medicine, 2014
- PET/MRI insert using digital SiPMs: Investigation of MR-compatibilityNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2014
- PETA4: a multi-channel TDC/ADC ASIC for SiPM readoutJournal of Instrumentation, 2013
- Performance of FBK SiPMs coupled to PETA3 read-out ASIC for PET applicationNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2013
- TOFPET ASIC for PET applicationsJournal of Instrumentation, 2013
- EASIROC, an Easy & Versatile ReadOut Device for SiPMPhysics Procedia, 2012
- ASIC development for SiPM readoutJournal of Instrumentation, 2009
- Positron Emission Tomography in Non–Small-Cell Lung Cancer: Prediction of Response to Chemotherapy by Quantitative Assessment of Glucose UseJournal of Clinical Oncology, 2003