Investigation of the Power Consumption of the PETsys TOFPET2 ASIC
Preprint
Abstract
In state-of-the-art positron emission (PET) tomography systems, application-specific integrated circuits (ASICs)are commonly used to precisely digitize the signals of analog silicon photo-multipliers (SiPMs). However, when operating PET electronics in a magnetic resonance (MR) system, one faces the challenge of mutual interference of these imaging techniques. To prevent signal deterioration along long analog signal lines, PET electronics with a low power consumption digitizing the signals close to the SiPMs are preferred. In this study, we evaluate the power consumption of the TOFPET2 ASIC. Its power consumption ranges from 3.6 to 7.2 mW/channel as a function of the input stage impedance and discriminator noise settings. We present an analytical model allowing to compute the power consumption of a given ASIC configuration. The configured input stage impedance and discriminator noise have an impact on the coincidence resolution time, energy resolution, and photon trigger level. Since the TOFPET2 ASIC delivers state-of-the-art performance with a power consumption similar or even lower than other ASICs typically used for PET applications, it is a favorable candidate to digitize the signals of SiPMs in future simultaneous PET/MR systems.Keywords
All Related Versions
This publication has 42 references indexed in Scilit:
- The current mode Time-over-Threshold ASIC for a MPPC module in a TOF-PET systemNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2018
- Experimental results with TOFPET2 ASIC for time-of-flight applicationsNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2018
- A compact, high-density gamma-detection module for Time-of-Flight measurements in PET applicationsNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2016
- Development of analog solid-state photo-detectors for Positron Emission TomographyNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2016
- Advances in time-of-flight PETPhysica Medica, 2016
- STIC3 – Silicon Photomultiplier Timing Chip with picosecond resolutionNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2015
- Time resolution deterioration with increasing crystal length in a TOF-PET systemNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2014
- PET/MRI insert using digital SiPMs: Investigation of MR-compatibilityNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2014
- Performance of FBK SiPMs coupled to PETA3 read-out ASIC for PET applicationNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2013
- EASIROC, an Easy & Versatile ReadOut Device for SiPMPhysics Procedia, 2012