A New Design for a Compact Centrifugal Blood Pump with a Magnetically Levitated Rotor
Authors:
- Tadahiko ShinshiTSTadahiko Shinshi
- Hideo HoshiHHHideo Hoshi
- Setsuo TakataniSTSetsuo Takatani
- Akira ShimokohbeASAkira Shimokohbe
- 1 November 2004
- journal article
- research article
- Published by Wolters Kluwer Health in ASAIO Journal
- Vol. 50 (6) , 550-556
Abstract
A compact centrifugal blood pump has been developed using a radial magnetic bearing with a two-degree of freedom active control. The proposed magnetic bearing exhibits high stiffness, even in passively controlled directions, and low power consumption because a permanent magnet, incorporated with the rotor, suspends its weight. The rotor is driven by a Lorentz force type of built-in motor, avoiding mechanical friction and material wear. The built-in motor is designed to generate only rotational torque, without radial and axial attractive forces on the rotor, leading to low power consumption by the magnetic bearing. The fabricated centrifugal pump measured 65 mm in diameter and 45 mm in height and weighed 0.36 kg. In the closed loop circuit filled with water, the pump provided a flow rate of 4.5 L/min at 2,400 rpm against a pressure head of 100 mm Hg. Total power consumption at that point was 18 W, including 2 W required for magnetic levitation, with a total efficiency of 5.7%. The experimental results showed that the design of the compact magnetic bearing was feasible and effective for use in a centrifugal blood pump.Keywords
This publication has 12 references indexed in Scilit:
- One Piece Ultracompact Totally Implantable Electromechanical Total Artificial Heart for Permanent UseASAIO Journal, 2002
- HeartMate III: Pump Design for a Centrifugal LVAD with a Magnetically Levitated RotorASAIO Journal, 2001
- Pressure-Volume Relationship of a Pulsatile Blood Pump for Ventricular Assist Device DevelopmentASAIO Journal, 2001
- Development Status of Terumo Implantable Left Ventricular Assist SystemArtificial Organs, 2001
- Development of a compact, sealless, tripod supported, magnetically driven centrifugal blood pump.Artificial Organs, 2000
- A novel compact PMSM with magnetic bearing for artificial heart applicationIEEE Transactions on Industry Applications, 2000
- New mechanism to reduce the size of the monopivot magnetic suspension blood pump: direct drive mechanism.Artificial Organs, 1997
- A Magnetically Suspended and Hydrostatically Stabilized Centrifugal Blood PumpArtificial Organs, 1996
- Prototype Continuous Flow Ventricular Assist Device Supported on Magnetic BearingsArtificial Organs, 1996
- Recent Studies of the Centrifugal Blood Pump with a Magnetically Suspended ImpellerArtificial Organs, 1995