An automated quasi-continuous capillary refill timing device
Open Access
- 7 December 2015
- journal article
- research article
- Published by IOP Publishing in Physiological Measurement
- Vol. 37 (1) , 83-99
Abstract
Capillary refill time (CRT) is a simple means of cardiovascular assessment which is widely used in clinical care. Currently, CRT is measured through manual assessment of the time taken for skin tone to return to normal colour following blanching of the skin surface. There is evidence to suggest that manually assessed CRT is subject to bias from ambient light conditions, a lack of standardisation of both blanching time and manually applied pressure, subjectiveness of return to normal colour, and variability in the manual assessment of time. We present a novel automated system for CRT measurement, incorporating three components: a non-invasive adhesive sensor incorporating a pneumatic actuator, a diffuse multi-wavelength reflectance measurement device, and a temperature sensor; a battery operated datalogger unit containing a self contained pneumatic supply; and PC based data analysis software for the extraction of refill time, patient skin surface temperature, and sensor signal quality. Through standardisation of the test, it is hoped that some of the shortcomings of manual CRT can be overcome. In addition, an automated system will facilitate easier integration of CRT into electronic record keeping and clinical monitoring or scoring systems, as well as reducing demands on clinicians. Summary analysis of volunteer (n = 30) automated CRT datasets are presented, from 15 healthy adults and 15 healthy children (aged from 5 to 15 years), as their arms were cooled from ambient temperature to 5°C. A more detailed analysis of two typical datasets is also presented, demonstrating that the response of automated CRT to cooling matches that of previously published studies.Keywords
Funding Information
- Medical Research Council (Developmental Pathway Funding Schem)
This publication has 23 references indexed in Scilit:
- Capillary refill time. Is it time to fill the gaps?Medical Journal Armed Forces India, 2013
- Capillary Refill TimeAnesthesia & Analgesia, 2011
- Impact of patient and environmental factors on capillary refill time in adultsThe American Journal of Emergency Medicine, 2008
- A Novel Imaging Technique to Measure Capillary-Refill Time: Improving Diagnostic Accuracy for Dehydration in Young Children With GastroenteritisPediatrics, 2006
- Clinical detection of low upper body blood flow in very premature infants using blood pressure, capillary refill time, and central-peripheral temperature differenceArchives of Disease in Childhood - Fetal and Neonatal Edition, 2004
- Defining normal capillary refill: Variation with age, sex, and temperatureAnnals of Emergency Medicine, 1988
- The validity and practicality of sun-reactive skin types I through VIArchives of Dermatology, 1988
- Trauma scoreCritical Care Medicine, 1981
- STUDIES ON GANGRENE FOLLOWING COLD INJURY: VI. CAPILLARY BLOOD FLOW AFTER COLD INJURY, THE EFFECTS OF RAPID WARMING, AND SYMPATHETIC BLOCK 1Journal of Clinical Investigation, 1947
- CYCLOPROPANE ANESTHESIAAnesthesiology, 1940