Conventional and phase contrast x-ray imaging techniques and ultrasound imaging method in breast tumor detection: initial comparison studies using phantom
Authors:
- 19 February 2018
- conference paper
- Published by SPIE-Intl Soc Optical Eng
- Vol. 10495, 38
Abstract
The objective of this study was to demonstrate the capability of the high-energy in-line phase contrast imaging in detecting the breast tumors which are undetectable by conventional x-ray imaging but detectable by ultrasound. Experimentally, a CIRS multipurpose breast phantom with heterogeneous 50% glandular and 50% adipose breast tissue was imaged by high-energy in-line phase contrast system, conventional x-ray system and ultrasonography machine. The high-energy in-line phase contrast projection was acquired at 120 kVp, 0.3 mAs with the focal spot size of 18.3 μm. The conventional x-ray projection was acquired at 40 kVp, 3.3 mAs with the focal spot size of 22.26 μm. Both of the x-ray imaging acquisitions were conducted with a unique mean glandular dose of 0.08 mGy. As the result, the high-energy in-line phase contrast system was able to detect one lesion-like object which was also detected by the ultrasonography. This object was spherical shape with the length of about 12.28 mm. Also, the conventional x-ray system was not able to detect any objects. This result indicated the advantages provided by high-energy in-line phase contrast over conventional x-ray system in detecting lesion-like object under the same radiation dose. To meet the needs of current clinical strategies for high-density breasts screening, breast phantoms with higher glandular densities will be employed in future studies.Keywords
This publication has 19 references indexed in Scilit:
- Detectability comparison between a high energy x-ray phase sensitive and mammography systems in imaging phantoms with varying glandular-adipose ratiosPhysics in Medicine & Biology, 2017
- Comparison of Tomosynthesis Plus Digital Mammography and Digital Mammography Alone for Breast Cancer ScreeningRadiology, 2013
- Breast Cancer Screening With Imaging: Recommendations From the Society of Breast Imaging and the ACR on the Use of Mammography, Breast MRI, Breast Ultrasound, and Other Technologies for the Detection of Clinically Occult Breast CancerJournal of the American College of Radiology, 2010
- Breast cancer detection using automated whole breast ultrasound and mammography in radiographically dense breastsEuropean Radiology, 2009
- Combined Screening With Ultrasound and Mammography vs Mammography Alone in Women at Elevated Risk of Breast CancerJAMA, 2008
- Phase retrieval and differential phase-contrast imaging with low-brilliance X-ray sourcesNature Physics, 2006
- Evaluation of edge effect due to phase contrast imaging for mammographyMedical Physics, 2005
- Clinical implementation of x‐ray phase‐contrast imaging: Theoretical foundations and design considerationsMedical Physics, 2003
- Low-dose phase contrast x-ray medical imagingPhysics in Medicine & Biology, 1998
- Contrast and resolution in imaging with a microfocus x-ray sourceReview of Scientific Instruments, 1997