Comparison of Actual Tidal Volume in Neonatal Lung Model Volume Control Ventilation using Three Ventilators
Authors:
Open Access
- 1 July 2010
- journal article
- research article
- Published by SAGE Publications in Anaesthesia and Intensive Care
- Vol. 38 (4) , 599-606
Abstract
In neonates, small changes in tidal volumes (V T ) may lead to complications. Previous studies have shown a significant difference between ventilator-measured tidal volume and tidal volume delivered (actual V T ). We evaluated the accuracy of three different ventilators to deliver small V T during volume-controlled ventilation. We tested Servo 300, 840 ventilator and Evita 4 Neoflow ventilators with lung models simulating normal and injured neonatal lung compliance models. Gas volume delivered from the ventilator into the test circuit (V TV ) and actual V T to the test lung were measured using Ventrak respiration monitors at set V T (30 ml). The gas volume increase of the breathing circuit was then calculated. Tidal volumes of the SV300 and PB840 in both lung models were similar to the set V T and the actual tidal volumes in the injured model (20.7 ml and 19.8 ml, respectively) were significantly less than that in the normal model (27.4 ml and 23.4 ml). PB840 with circuit compliance compensation could not improve the actual V T . V TV of the EV4N in the normal and the injured models (37.8 ml and 46.6 ml) were markedly increased compared with set V T , and actual V T were similar to set V T in the normal and injured model (30.2 ml and 31.9 ml, respectively). EV4N measuring V T close to the lung could match actual V T to almost the same value as the set V T , however the gas volume of the breathing circuit was increased. If an accurate value for the patient's actual V T is needed, this V T must be measured by a sensor located between the Y-piece and the tracheal tube.Keywords
This publication has 13 references indexed in Scilit:
- Bench test evaluation of volume delivered by modern ICU ventilators during volume-controlled ventilationIntensive Care Medicine, 2010
- Growth and Development2009
- Volume-targeted ventilationEarly Human Development, 2006
- Circuit compliance compensation in lung protective ventilation2006
- The influence of flow rate on breathing circuit compliance and tidal volume delivered to patients in mechanical ventilationPhysiological Measurement, 2005
- Effect of lung compliance and endotracheal tube leakage on measurement of tidal volumeCritical Care, 2004
- Tidal Volumes For Ventilated Infants Should Be Determined with a Pneumotachometer Placed at the Endotracheal TubeAmerican Journal of Respiratory and Critical Care Medicine, 2000
- Ventilation with Lower Tidal Volumes as Compared with Traditional Tidal Volumes for Acute Lung Injury and the Acute Respiratory Distress SyndromeNew England Journal of Medicine, 2000
- Lung Protective Strategies of Ventilation in the Neonate: What Are They?Pediatrics, 2000
- Ventilator-induced Lung Injury: Lessons from Experimental StudiesAmerican Journal of Respiratory and Critical Care Medicine, 1998