Abstract
An integrated detection system, consists of a thin diamond sensor incorporated into a compact signal loop, coupled in turn to an application-specific integrated circuit readout chip, has been tested for the first time, using ps-duration electron pulses from the Next Linear Collider Test Accelerator at the SLAC National Accelerator Laboratory. For readout frame rates as high as 120 kHz, the detection system is shown to achieve a linear dynamic range of 10 3 and a bandwidth of 4–5 GHz, with no ringing seen on the tail of the output pulse. These results represent the best-performing multi-GHz ionizing particle detection system achieved to date.
Funding Information
  • U.S. Department of Energy (89233218CNA000001)
  • Basic Energy Sciences (DE-SC0024205)
  • Basic Energy Sciences (DE-AC02 05CH11521)
  • Basic Energy Sciences (A24-0496-S001)
  • Basic Energy Sciences (DE-AC02-76SF00515)
  • High Energy Physics (DE-SC0010107)
  • University of California, Santa Cruz (LFR-20-653232)
  • Sandia National Laboratories (DE-NA-0003525)
  • FWP (LANLE8AN)

This publication has 9 references indexed in Scilit: