Solid-state pulse modulator for a 1.7-MW X-band magnetron
Authors:
- 22 May 2014
- journal article
- Published by Korean Physical Society in Journal of the Korean Physical Society
- Vol. 64 (9) , 1267-1271
Abstract
Medical linear accelerators (LINAC) for cancer treatment require pulse modulators to generate high-power pulses with a fast rise time, flat top and short duration to drive high-power magnetrons. Solid-state pulse modulators (SSPM) for medical LINACs that use high power semiconductor switches with high repetition rates, high stability and long lifetimes have been introduced to replace conventional linear-type pulse generators that use gaseous discharge switches. In this paper, the performance of a developed SSPM, which mainly consists of a capacitor charger, an insulatedgate bipolar transistor (IGBT)–capacitor stack and a pulse transformer, is evaluated with a dummy load and an X-band magnetron load. A theoretical analysis of the pulse transformer, which is a critical element of the SSPM, is carried out. The output pulse has a fast rise time and low droop, such that the modulator can drive the X-band magnetron.Keywords
This publication has 3 references indexed in Scilit:
- Portable X-Band Linear Electron Accelerators for Radiographic Applications2006
- Portable, X-Band, Linear Accelerator SystemsIEEE Transactions on Nuclear Science, 1985
- Designing pulse transformers for small sizeIEEE Transactions on Magnetics, 1977