Jaegu Choi
Current Organizations / Institutes
The organizations/institutes from the scholar's latest article
Jaegu Choi has a total of 10 co-authors. Here are the top co-authors by number of shared publications:
Dong-Goo Kang
Korea Electrotechnology Research Institute, Changwon, Gyeongsangnam-do, Korea (the Republic of)
2 shared publicationsKyuseok Kim
Korea Electrotechnology Research Institute, Changwon, Gyeongsangnam-do, Korea (the Republic of)
2 shared publicationsYoung-Wook Choi
Korea Electrotechnology Research Institute (KERI)
2 shared publicationsKwan-Ho Kim
Korea Electrotechnology Research Institute (KERI)
1 shared publication- Kyuseok Kim
Electro-Medical Device Research Center, Korea Electrotechnology Research Institute (KERI), 111, Hanggaul-ro, Sangnok-gu, Ansan-si, Gyeonggi-do 15588, Korea
1 shared publication Sun-Jo Hwang
Korea Electrotechnology Research Institute
1 shared publication
Latest Publications
- 14 October 2020
- Published by IOP Publishing in Japanese Journal of Applied Physics
- Vol. 59 (11) , 115001
Thin-film emitter based on a tungsten-DLC multilayer
Nanostructure emitters are an alternative low-power approach to generate electrons and provide control of the electron energy, digital control, and a small device size compared to that of conventional thermal emitters. Carbon nanotubes (CNTs) have been intensively investigated and reported to have excellent properties due to their high aspect ratio, high thermal conductivity, and chemical inertness. However, manufactured CNTs have uniformity issues, and this leads to problems with the characterization of X-ray tubes, including their current and lifetime. In this study, we propose a new cold cathode emitter based on a tungsten and diamond-like carbon multilayer with a sapphire supported plate to ensure durability and a homogeneous manufactured structure. We conducted the deposition of the emitter components at the nanoscale by using a homemade high vacuum system to demonstrate its viability for X-ray sources. A numerical simulation was also conducted to demonstrate its viability.
- 14 October 2020
- Published by IOP Publishing in Japanese Journal of Applied Physics
- Vol. 59 (11) , 115001
Thin-film emitter based on a tungsten-DLC multilayer
Nanostructure emitters are an alternative low-power approach to generate electrons and provide control of the electron energy, digital control, and a small device size compared to that of conventional thermal emitters. Carbon nanotubes (CNTs) have been intensively investigated and reported to have excellent properties due to their high aspect ratio, high thermal conductivity, and chemical inertness. However, manufactured CNTs have uniformity issues, and this leads to problems with the characterization of X-ray tubes, including their current and lifetime. In this study, we propose a new cold cathode emitter based on a tungsten and diamond-like carbon multilayer with a sapphire supported plate to ensure durability and a homogeneous manufactured structure. We conducted the deposition of the emitter components at the nanoscale by using a homemade high vacuum system to demonstrate its viability for X-ray sources. A numerical simulation was also conducted to demonstrate its viability.
- Open Access
Effectiveness of Non-Local Means Algorithm with an Industrial 3 MeV LINAC High-Energy X-ray System for Non-Destructive Testing
Industrial high-energy X-ray imaging systems are widely used for non-destructive testing (NDT) to detect defects in the internal structure of objects. Research on X-ray image noise reduction techniques using image processing has been widely conducted with the aim of improving the detection of defects in objects. In this paper, we propose a non-local means (NLM) denoising algorithm to improve the quality of images obtained using an industrial 3 MeV high-energy X-ray imaging system. We acquired X-ray images using various castings and assessed the performance visually and by obtaining the intensity profile, contrast-to-noise ratio, coefficient of variation, and normalized noise power spectrum. Overall, the quality of images processed by the proposed NLM algorithm is superior to those processed by existing algorithms for the acquired casting images. In conclusion, the NLM denoising algorithm offers an efficient and competitive approach to overcome the noise problem in high-energy X-ray imaging systems, and we expect the accompanying image processing software to facilitate and improve image restoration.
Open AccessEffectiveness of Non-Local Means Algorithm with an Industrial 3 MeV LINAC High-Energy X-ray System for Non-Destructive Testing
Industrial high-energy X-ray imaging systems are widely used for non-destructive testing (NDT) to detect defects in the internal structure of objects. Research on X-ray image noise reduction techniques using image processing has been widely conducted with the aim of improving the detection of defects in objects. In this paper, we propose a non-local means (NLM) denoising algorithm to improve the quality of images obtained using an industrial 3 MeV high-energy X-ray imaging system. We acquired X-ray images using various castings and assessed the performance visually and by obtaining the intensity profile, contrast-to-noise ratio, coefficient of variation, and normalized noise power spectrum. Overall, the quality of images processed by the proposed NLM algorithm is superior to those processed by existing algorithms for the acquired casting images. In conclusion, the NLM denoising algorithm offers an efficient and competitive approach to overcome the noise problem in high-energy X-ray imaging systems, and we expect the accompanying image processing software to facilitate and improve image restoration.
- 22 May 2014
- Published by Korean Physical Society in Journal of the Korean Physical Society
- Vol. 64 (9) , 1267-1271
Solid-state pulse modulator for a 1.7-MW X-band magnetron
Medical linear accelerators (LINAC) for cancer treatment require pulse modulators to generate high-power pulses with a fast rise time, flat top and short duration to drive high-power magnetrons. Solid-state pulse modulators (SSPM) for medical LINACs that use high power semiconductor switches with high repetition rates, high stability and long lifetimes have been introduced to replace conventional linear-type pulse generators that use gaseous discharge switches. In this paper, the performance of a developed SSPM, which mainly consists of a capacitor charger, an insulatedgate bipolar transistor (IGBT)–capacitor stack and a pulse transformer, is evaluated with a dummy load and an X-band magnetron load. A theoretical analysis of the pulse transformer, which is a critical element of the SSPM, is carried out. The output pulse has a fast rise time and low droop, such that the modulator can drive the X-band magnetron.
- 22 May 2014
- Published by Korean Physical Society in Journal of the Korean Physical Society
- Vol. 64 (9) , 1267-1271
Solid-state pulse modulator for a 1.7-MW X-band magnetron
Medical linear accelerators (LINAC) for cancer treatment require pulse modulators to generate high-power pulses with a fast rise time, flat top and short duration to drive high-power magnetrons. Solid-state pulse modulators (SSPM) for medical LINACs that use high power semiconductor switches with high repetition rates, high stability and long lifetimes have been introduced to replace conventional linear-type pulse generators that use gaseous discharge switches. In this paper, the performance of a developed SSPM, which mainly consists of a capacitor charger, an insulatedgate bipolar transistor (IGBT)–capacitor stack and a pulse transformer, is evaluated with a dummy load and an X-band magnetron load. A theoretical analysis of the pulse transformer, which is a critical element of the SSPM, is carried out. The output pulse has a fast rise time and low droop, such that the modulator can drive the X-band magnetron.
- 27 May 2012
- Published by Korean Physical Society in Journal of the Korean Physical Society
- Vol. 60 (9) , 1457-1463
Focal spot calibration in a digital breast tomosynthesis system
- 27 May 2012
- Published by Korean Physical Society in Journal of the Korean Physical Society
- Vol. 60 (9) , 1457-1463
Focal spot calibration in a digital breast tomosynthesis system