Modular Hip Implant Fracture at the Stem-Sleeve Interface
Authors:
- Thomas ParisiTPThomas Parisi
- Brian BurroughsBBBrian Burroughs
- Young-Min KwonYKYoung-Min Kwon
- 1 March 2015
- journal article
- case report
- Published by SLACK, Inc. in Orthopedics
- Vol. 38 (3) , e234-e239
Abstract
The use of modular implants in femoral stem design has grown increasingly popular over the last decade because of the theoretical advantage of more flexibility and optimization of femoral anteversion, limb length, and femoral component offset. With the benefit of increased surgical flexibility, however, modularity also carries the theoretical risks of fretting at the modular surfaces, sequelae of wear debris, and possible failure and fracture of the stem at the modular junction. Indeed, there have been an increasing number of reports of modular implants failing due to fracture at modular junctions. The S-ROM prosthesis (DePuy Orthopaedics, Inc, Warsaw, Indiana), however, has a stellar clinical record and has been used with good results in both primary and revision total hip arthroplasty. Only a single case of S-ROM failure at the stem-sleeve interface has been reported in the orthopedic literature. The aim of this case report was to present a succinct history of proximal modularity in total hip arthroplasty and to describe the only known case of this type of catastrophic failure in an S-ROM prosthesis with a metal-on-metal bearing. Despite a low level of serum metal ions on presentation, scanning electron microscopy showed findings consistent with corrosive processes and pseudotumor was seen at revision surgery. [ Orthopedics. 2015; 38(3):e234–e239.]Keywords
This publication has 19 references indexed in Scilit:
- Failure of a Modular Hip Implant at the Stem-Sleeve InterfaceOrthopedics, 2013
- Total Hip Arthroplasty Modular Neck FailureThe Journal of Arthroplasty, 2011
- Failure of the Modular Neck in a Total Hip ArthroplastyThe Journal of Arthroplasty, 2010
- Catastrophic failure of 4 cobalt-alloy Omnifit hip arthroplasty femoral componentsActa Orthopaedica, 2008
- Fretting wear in a modular neck hip prosthesisJournal of Biomedical Materials Research, 1997
- Design-related fretting wear in modular neck hip prosthesisJournal of Biomedical Materials Research, 1996
- Torsional resistance and wear of a modular sleeve-stem hip systemClinical Materials, 1993
- Fracture of hip prostheses due to inadequate weldingActa Orthopaedica, 1987
- Fracture of Femoral Prostheses in Total Hip ReplacementClinical Orthopaedics and Related Research, 1975
- Factors in the Mechanical Failure of the Femoral Component in Total Hip Prosthesis:Report of Six Fatigue Fractures of the Femoral Stern and Results of Experimental Loading TestsActa Orthopaedica, 1974