A Report on the Early Failure of the LPM Proximal Interphalangeal Joint Replacement
Authors:
- J L HobbyJHJ L Hobby
- Research and Audit Committee of the BSSHRBResearch and Audit Committee of the BSSH
Open Access
- 1 August 2008
- journal article
- research article
- Published by SAGE Publications in Journal of Hand Surgery (European Volume)
- Vol. 33 (4) , 526-527
Abstract
The LPM (Van Straten Medical, Netherlands) proximal interphalangeal joint replacement was introduced to the United Kingdom in October 2000, and 257 implants were sold. We have audited the results of this prosthesis, and obtained data for 164 of the 257 implants. Forty-seven (29%) of the 164 have been revised and another 33 (20%) are showing clinical and radiological signs of failure at a maximum follow-up of 6 years. Massive osteolysis leading to aseptic loosening is the commonest cause of failure. We believe that all patients who have had this prosthesis inserted should be kept under regular clinical and radiographic review.Keywords
This publication has 7 references indexed in Scilit:
- TWO TO FIVE YEAR FOLLOW-UP OF THE LPM CERAMIC COATED PROXIMAL INTERPHALANGEAL JOINT ARTHROPLASTYJournal of Hand Surgery (European Volume), 2008
- ReplyJournal of Hand Surgery (European Volume), 2006
- Vicryl Rapide® Versus Vicryl® Suture in Skin Closure of the Hand in Children: A Randomized Prospective StudyJournal of Hand Surgery (European Volume), 2005
- Challenges in Evaluating Patients Lost to Follow-up in Clinical Studies of Rotator Cuff Tears*Journal of Bone and Joint Surgery, 2000
- Loss To Follow-up MattersThe Journal of Bone and Joint Surgery. British volume, 1997
- An Arthroscopic Technique for Anterior Stabilization of the Shoulder with a Bioabsorbable Tack*Journal of Bone and Joint Surgery, 1996
- Proximal interphalangeal joint silicone replacement arthroplasty: Clinical results using an anterior approachThe Journal of Hand Surgery, 1995