Platelet Adhesion Dynamics Following High Mechanical Shear Stress Exposure: Distinct Mechanisms and a Unified Kinetic Framework

Authors:

Abstract
Background: Platelet adhesion to collagen and von Willebrand factor (VWF) is essential for primary hemostasis. High mechanical shear stress (HMSS) generated by blood pumps induces blood damage that is strongly associated with hemostatic disorders in patients receiving mechanical circulatory support (MCS). However, the dynamic process of platelet adhesion and thrombus formation following HMSS exposure remains poorly understood.Methods: Fresh healthy human blood was circulated in an in vitro blood pump loop under clinically relevant operating conditions. Baseline blood and HMSS-exposed blood collected after 2 and 4 h were perfused through collagen- or VWF-coated microfluidic channels for real-time visualization and quantitative analysis of platelet adhesion. Platelet activation, platelet receptor expression, and VWF degradation were also evaluated.Results: On collagen-coated surfaces, normal platelets readily adhered and progressively formed enlarging, stable thrombi. In contrast, HMSS-exposed platelets formed predominantly smaller aggregates and showed a progressive loss of large thrombi. On VWF-coated surfaces, platelets adhered predominantly as individual cells, while HMSS exposure increased platelet translocation, indicating reduced adhesion stability. Despite these distinct adhesion phenotypes and different degrees of HMSS-induced blood damage, the temporal evolution of platelet accumulation under all experimental conditions was consistently described by the same first-order saturation relationship.Conclusions: HMSS-induced blood damage differentially impairs platelet adhesive function by limiting thrombus propagation on collagen while weakening platelet–VWF interactions. A simple kinetic framework is introduced to describe the temporal evolution of platelet accumulation, providing a quantitative tool for evaluating platelet adhesive function and the hemocompatibility of MCS devices.